Correcting heat‐induced chemical shift distortions in proton resonance frequency‐shift thermometry
- 24 August 2015
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 76 (1) , 172-182
Abstract
Purpose: To reconstruct proton resonance frequency‐shift temperature maps free of chemical shift distortions.Theory and Methods: Tissue heating created by thermal therapies such as focused ultrasound surgery results in a change in proton resonance frequency that causes geometric distortions in the image and calculated temperature maps, in the same manner as other chemical shift and off‐resonance distortions if left uncorrected. We propose an online‐compatible algorithm to correct these distortions in 2DFT and echo‐planar imaging acquisitions, which is based on a k‐space signal model that accounts for proton resonance frequency change‐induced phase shifts both up to and during the readout. The method was evaluated with simulations, gel phantoms, and in vivo temperature maps from brain, soft tissue tumor, and uterine fibroid focused ultrasound surgery treatments.Results: Without chemical shift correction, peak temperature and thermal dose measurements were spatially offset by approximately 1 mm in vivo. Spatial shifts increased as readout bandwidth decreased, as shown by up to 4‐fold greater temperature hot spot asymmetry in uncorrected temperature maps. In most cases, the computation times to correct maps at peak heat were less than 10 ms, without parallelization.Conclusion: Heat‐induced proton resonance frequency changes create chemical shift distortions in temperature maps resulting from MR‐guided focused ultrasound surgery ablations, but the distortions can be corrected using an online‐compatible algorithm. Magn Reson Med 76:172–182, 2016.Keywords
Funding Information
- National Institutes of Health (R25CA136440, T32EB014841, P01CA159992)
- Focused Ultrasound Foundation
- Howard Hughes Medical Institute/Vanderbilt University Medical Center Certificate Program in Molecular Medicine
This publication has 27 references indexed in Scilit:
- Correction of proton resonance frequency shift MR‐thermometry errors caused by heat‐induced magnetic susceptibility changes during high intensity focused ultrasound ablations in tissues containing fatMagnetic Resonance in Medicine, 2013
- A Pilot Study of Focused Ultrasound Thalamotomy for Essential TremorNew England Journal of Medicine, 2013
- Hybrid proton resonance frequency/T1 technique for simultaneous temperature monitoring in adipose and aqueous tissuesMagnetic Resonance in Medicine, 2012
- Reconstruction of fully three‐dimensional high spatial and temporal resolution MR temperature maps for retrospective applicationsMagnetic Resonance in Medicine, 2011
- Transcranial Magnetic Resonance Imaging– Guided Focused Ultrasound Surgery of Brain TumorsNeurosurgery, 2010
- MR thermometryJournal of Magnetic Resonance Imaging, 2008
- MR thermometry for monitoring tumor ablationEuropean Radiology, 2007
- Magnetic resonance temperature imaging for guidance of thermotherapyJournal of Magnetic Resonance Imaging, 2000
- Noninvasive MRI thermometry with the proton resonance frequency method: Study of susceptibility effectsMagnetic Resonance in Medicine, 1995
- Correction for geometric distortion in echo planar images from B0 field variationsMagnetic Resonance in Medicine, 1995