Abstract
Single cell microfluidic impedance cytometry has been used to identify cells at high speed, on the basis of their dielectric properties. However, there is no electrical analogue to a fluorescent label, meaning that it is not possible to identify subpopulations of cells. We demonstrate discrimination and enumeration of antigenically defined cell subpopulations using an alternating current (AC) impedance labeling method. Small antibody conjugated beads are mixed with cells and bind to the target population, changing the electrical properties of the target subset of cells. The principle of the technique is demonstrated by identifying and enumerating the CD4 T-lymphocyte subpopulation in human whole blood. The technique represents a simple method for detecting a subpopulation of cells within a heterogeneous mix. The impedance-based antibody identification method could form the basis of simple low-cost point of care diagnostic technologies.Keywords
This publication has 19 references indexed in Scilit:
- Leukocyte analysis and differentiation using high speed microfluidic single cell impedance cytometryLab on a Chip, 2009
- Measurement of CD4+T cells in point-of-care settings with the Sysmex pocH-100i haematological analyserInternational Journal of Laboratory Hematology, 2009
- Enhancing the performance of a point-of-care CD4+ T-cell counting microchip through monocyte depletion for HIV/AIDS diagnosticsLab on a Chip, 2009
- Affordable CD4 T-cell enumeration for resource-limited regions: A status report for 2008Cytometry Part B: Clinical Cytometry, 2008
- A microfluidic device for practical label-free CD4+ T cell counting of HIV-infected subjectsLab on a Chip, 2006
- Deterministic hydrodynamics: Taking blood apartProceedings of the National Academy of Sciences, 2006
- Use of total lymphocyte count for informing when to start antiretroviral therapy in HIV-infected children: a meta-analysis of longitudinal dataThe Lancet, 2005
- A Microchip CD4 Counting Method for HIV Monitoring in Resource-Poor SettingsPLOS Medicine, 2005
- Determination of CD4 antigen density on cells: Role of antibody valency, avidity, clones, and conjugationCytometry, 1998
- Pneumocystis cariniiPneumonia and Mucosal Candidiasis in Previously Healthy Homosexual MenNew England Journal of Medicine, 1981