Diagnostic Accuracy of Urine Reagent Strips for Rapid Bacterial Meningitis Diagnosis: A Cross-Sectional Validation Study in a Resource-Limited Setting

Authors:

Abstract
Objective: To prospectively confirm the diagnostic performance of urine reagent strips in the detection of meningitis. Methodology: This cross-sectional study included patients aged 20–60 years with clinically suspected meningitis requiring lumbar puncture. The index test was a standardized urine reagent strip (Combur-10 Test®) applied to undiluted CSF, interpreted for leukocyte esterase, protein, and glucose according to manufacturer color gradings taking laboratory CSF analysis as gold standard test. Following parameters were statistically calculated for accuracy. Results: Among 105 enrolled patients (mean age 38.5 ± 12.2 years; 55.2% male), the prevalence of meningitis by reference standard was 60.0% (63/105). Against this gold standard, the reagent strip method demonstrated a sensitivity of 86.5% (95% CI: 75.8–93.1), specificity of 88.4% (95% CI: 74.1–95.6), positive predictive value of 91.5% (95% CI: 82.0–96.2), and negative predictive value of 82.6% (95% CI: 69.7–90.8). The diagnostic accuracy is 87.6% (95% CI: 79.9 – 92.9). The area under the ROC curve was 0.91 (95% CI: 0.86 – 0.96) which shows excellent discriminatory ability. Conclusion: Urine reagent strip are easily available, are cost effective and operationally feasible bedside diagnostic kits which can be used for a rapid diagnosis of meningitis in resource-limited settings.