Nasal masks or binasal prongs for delivering continuous positive airway pressure in preterm neonates—a randomised trial
Authors:
- 13 January 2017
- journal article
- research article
- Published by Springer Nature in European Journal of Pediatrics
- Vol. 176 (3) , 379-386
Abstract
The objective of this study was to compare the efficacy and safety of continuous positive airway pressure (CPAP) delivered using nasal masks with binasal prongs. We randomly allocated 72 neonates between 26 and 32 weeks gestation to receive bubble CPAP by either nasal mask (n = 37) or short binasal prongs (n = 35). Primary outcome was mean FiO2 requirement at 6, 12 and 24 h of CPAP initiation and the area under curve (AUC) of FiO2 against time during the first 24 h (FiO2 AUC0-24). Secondary outcomes were the incidence of CPAP failure and nasal trauma. FiO2 requirement at 6, 12 and 24 h (mean (SD); 25 (5.8) vs. 27.9 (8); 23.8 (4.5) vs. 25.4 (6.8) and 22.6 (6.8) vs. 22.7 (3.3)) as well as FiO2 AUC0-24 (584.0 (117.8) vs. 610.6 (123.6)) were similar between the groups. There was no difference in the incidence of CPAP failure (14 vs. 20%; relative risk 0.67; 95% confidence interval 0.24-1.93). Incidence of severe nasal trauma was lower with the use of nasal masks (0 vs. 31%; p < .001). Nasal masks appear to be as efficacious as binasal prongs in providing CPAP. Masks are associated with lower risk of severe nasal trauma. CTRI2012/08/002868 What is Known? • Binasal prongs are better than single nasal and nasopharyngeal prongs for delivering continuous positive airway pressure (CPAP) in preventing need for re-intubation. • It is unclear if they are superior to newer generation nasal masks in preterm neonates requiring CPAP. What is New? • Oxygen requirement during the first 24 h of CPAP delivery is comparable with use of nasal masks and binasal prongs. • Use of nasal masks is, however, associated with significantly lower risk of severe grades of nasal injury.Keywords
This publication has 27 references indexed in Scilit:
- Randomized Trial of Prongs or Mask for Nasal Continuous Positive Airway Pressure in Preterm InfantsPediatrics, 2012
- Survey of nasal continuous positive airways pressure (NCPAP) and nasal intermittent positive pressure ventilation (NIPPV) use in Irish newborn nurseries: Table 1Archives of Disease in Childhood - Fetal and Neonatal Edition, 2010
- Nasal trauma due to continuous positive airway pressure in neonatesArchives of Disease in Childhood - Fetal and Neonatal Edition, 2010
- Clinical Prediction Score for Nasal CPAP Failure in Pre-term VLBW Neonates with Early Onset Respiratory DistressJournal of Tropical Pediatrics, 2010
- Early CPAP versus Surfactant in Extremely Preterm InfantsNew England Journal of Medicine, 2010
- Devices and pressure sources for administration of nasal continuous positive airway pressure (NCPAP) in preterm neonatesCochrane Database of Systematic Reviews, 2008
- National Pressure Ulcer Advisory Panel's Updated Pressure Ulcer Staging SystemAdvances in Skin & Wound Care, 2007
- Incidence of nasal trauma associated with nasal prong versus nasal mask during continuous positive airway pressure treatment in very low birthweight infants: a randomised control studyArchives of Disease in Childhood - Fetal and Neonatal Edition, 2005
- Brief report. Predictors of failure of nasal continuous positive airway pressure in treatment of preterm infants with respiratory distress syndromeJournal of Tropical Pediatrics, 2000
- Nasal deformities resulting from flow driver continuous positive airway pressure.Archives of Disease in Childhood - Fetal and Neonatal Edition, 1996