Investigations into the Geometric Calibration and Systematic Effects of a Micro-CT System
Open Access
Abstract
Micro-Computed Tomography (µCT) systems are used for examining the internal structures of various objects, such as material samples, manufactured parts, and natural objects. Resolving fine details or performing accurate geometric measurements in the voxel data critically depends on the precise calibration of the µCT systems geometry. This paper presents a calibration method for µCT systems using projections of a calibration phantom, where the coordinates of the phantom are initially unknown. The approach involves detecting and tracking steel ball bearings and adjusting the unknown system geometry parameters using non-linear least squares optimization. Multiple geometric models are tested to verify their suitability for a self-calibration approach. The implementation is tested using a calibration phantom captured at different magnifications. The results demonstrate the system’s capability to determine the geometry model parameters with a remaining error on the detector between 0.27 px and 0.18 px. Systematic errors that remain after calibration, as well as changing parameters due to system instabilities, are investigated. The source code of this work is published to enable further research.Keywords
Funding Information
- Deutsche Forschungsgemeinschaft (417002380)
This publication has 35 references indexed in Scilit:
- Investigating focal spot position drift in a mobile imaging system equipped with a monobloc‐based x‐ray generatorMedical Physics, 2023
- Advances in the metrological traceability and performance of X-ray computed tomographyCIRP Annals, 2022
- Calibration and rectification of bi-telecentric lenses in Scheimpflug conditionOptics and Lasers in Engineering, 2022
- Knowledge-based self-calibration method of calibration phantom by and for accurate robot-based CT imaging systemsKnowledge-Based Systems, 2021
- Robust detection of non-overlapping ellipses from points with applications to circular target extraction in images and cylinder detection in point cloudsISPRS Journal of Photogrammetry and Remote Sensing, 2021
- Measurement of sample stage error motions in cone-beam X-ray computed tomography instruments by minimization of reprojection errorsPrecision Engineering, 2021
- Software-based compensation of instrument misalignments for X-ray computed tomography dimensional metrologyPrecision Engineering, 2018
- Measurement of the X-ray computed tomography instrument geometry by minimization of reprojection errors—Implementation on experimental dataPrecision Engineering, 2018
- Evaluating the effects of detector angular misalignments on simulated computed tomography dataPrecision Engineering, 2016
- Sensor modelling and camera calibration for close-range photogrammetryISPRS Journal of Photogrammetry and Remote Sensing, 2016