Abstract
Abstract:Accurate imaging of impedance changes in the brain with EIT using scalp electrodes ideally requires an algorithm designed for a 3D hemispherical object that is capable of imaging through the skull. A 2D algorithm is presented, intended as the first step towards a full 3D version. It is based on a sensitivity matrix approach and allows images to be reconstructed from any electrode positions. Its performance was assessed using a 2D circular tank with a simulated skull. The findings suggested the following: if polar current drive was used with no explicit skull compensation, a feature in the image would be placed 25% closer to the center and would be around one‐third the amplitude of an identical perturbation in a homogeneous medium.

This publication has 14 references indexed in Scilit: