Comparison of the Carpentier-Edwards Perimount™ and St. Jude Medical Epic™ Bioprostheses for Aortic Valve Replacement-A Retrospective Echocardiographic Short-Term Study
- 1 May 2009
- journal article
- research article
- Published by Hindawi Limited in Journal of Cardiac Surgery
- Vol. 24 (3) , 260-264
Abstract
Stented biological prostheses show different performance related to their design and ultrastructure. Therefore, we compared the Carpentier-Edwards Perimount (Edwards Lifesciences, Irvine, CA, USA) and the St. Jude Medical Epic (SJM Epic; St. Jude Medical Inc., St. Paul, MN, USA) prostheses with respect to their hemodynamic features. Eighty-six patients underwent aortic valve replacement for severe aortic stenosis (50 patients), aortic regurgitation (seven patients), or mixed lesion (29 patients) with these randomly assigned bioprostheses (Perimount: 39 patients; SJM Epic: 47 patients). Following informed consent, patients were examined postoperatively (mean 9.7 +/- 4 months after surgery) with a detailed echocardiographic study, quality of life was determined by a standardized questionnaire. Comparing the indexed effective orifice area (0.77 cm(2)/m(2) vs. 0.64 cm(2)/m(2); p = 0.019), the mean pressure gradient (11.15 mmHg vs. 14.80 mmHg; p = 0.004) and the stroke-work loss (6.95% vs. 8.49%; p = 0.025), the Carpentier-Edwards Perimount demonstrated significantly better hemodynamic features than the SJM Epic prosthesis. These effects were more pronounced in the smaller prosthetic sizes. The Carpentier-Edwards Perimount prosthesis was less obstructive than the SJM Epic in echocardiographic controls. Patient-prosthesis mismatch as determined by indexed effective orifice area showed no impact on functional recovery.Keywords
This publication has 21 references indexed in Scilit:
- Doppler echocardiography markedly underestimates left ventricular stroke work loss in severe aortic valve stenosisEuropean Heart Journal - Cardiovascular Imaging, 2007
- Influence of completely supra-annular placement of bioprostheses on exercise hemodynamics in patients with a small aortic annulusThe Journal of Thoracic and Cardiovascular Surgery, 2007
- Patient prosthesis mismatch affects short- and long-term outcomes after aortic valve replacement☆European Journal of Cardio-Thoracic Surgery, 2006
- Small aortic annulus: The hydrodynamic performances of 5 commercially available tissue valvesThe Journal of Thoracic and Cardiovascular Surgery, 2006
- Impact of Prosthesis–Patient Size on Functional Recovery After Aortic Valve ReplacementCirculation, 2005
- Choice of prosthetic heart valve for adult patientsJACC, 2003
- Prosthesis-Patient Mismatch Affects Survival After Aortic Valve ReplacementCirculation, 2000
- Hemodynamic and clinical impact of prosthesis–patient mismatch in the aortic valve position and its preventionJACC, 2000
- Aortic valve replacement: Is valve size important?The Journal of Thoracic and Cardiovascular Surgery, 2000
- The problem of valve prosthesis-patient mismatch.Circulation, 1978