Real-Time Electrical Bioimpedance Characterization of Neointimal Tissue for Stent Applications
Open Access
Abstract
To follow up the restenosis in arteries stented during an angioplasty is an important current clinical problem. A new approach to monitor the growth of neointimal tissue inside the stent is proposed on the basis of electrical impedance spectroscopy (EIS) sensors and the oscillation-based test (OBT) circuit technique. A mathematical model was developed to analytically describe the histological composition of the neointima, employing its conductivity and permittivity data. The bioimpedance model was validated against a finite element analysis (FEA) using COMSOL Multiphysics software. A satisfactory correlation between the analytical model and FEA simulation was achieved in most cases, detecting some deviations introduced by the thin “double layer” that separates the neointima and the blood. It is hereby shown how to apply conformal transformations to obtain bioimpedance electrical models for stack-layered tissues over coplanar electrodes. Particularly, this can be applied to characterize the neointima in real-time. This technique is either suitable as a main mechanism for restenosis follow-up or it can be combined with proposed intelligent stents for blood pressure measurements to auto-calibrate the sensibility loss caused by the adherence of the tissue on the micro-electro-mechanical sensors (MEMSs).Keywords
This publication has 22 references indexed in Scilit:
- The Bio-Oscillator: A Circuit for Cell-Culture AssaysIEEE Transactions on Circuits and Systems II: Express Briefs, 2014
- Implantable Impedance PlethysmographySensors, 2014
- In-Stent NeoatherosclerosisJACC, 2012
- Extension to the analytical model of the interdigital electrodes capacitance for a multi-layered structureSensors and Actuators A: Physical, 2011
- A novel impedance-based tomography approach for stenotic plaque detection: A simulation studyInternational Journal of Cardiology, 2010
- Towards a self-reporting coronary artery stent—Measuring neointimal growth associated with in-stent restenosis using electrical impedance techniquesBiosensors and Bioelectronics, 2010
- Comparison of a Polymer-Based Paclitaxel-Eluting Stent With a Bare Metal Stent in Patients With Complex Coronary Artery DiseaseJAMA, 2005
- Revisiting the partial-capacitance approach to the analysis of coplanar transmission lines on multilayered substratesIEEE Transactions on Microwave Theory and Techniques, 2003
- Interpretation and prediction of triple-layer model capacitances and the structure of the oxide-electrolyte-water interfaceGeochimica et Cosmochimica Acta, 2001
- Four-Electrode Null Techniques for Impedance Measurement with High ResolutionReview of Scientific Instruments, 1968