Dependency of EBT2 film calibration curve on postirradiation time
Authors:
- 27 January 2014
- journal article
- Published by Wiley in Medical Physics
- Vol. 41 (2) , 021726
Abstract
Purpose:: The Ashland Inc. product EBT2 film model is a widely used quality assurance tool, especially for verification of 2‐dimensional dose distributions. In general, the calibration film and the dose measurement film are irradiated, scanned, and calibrated at the same postirradiation time (PIT), 1‐2 days after the films are irradiated. However, for a busy clinic or in some special situations, the PIT for the dose measurement film may be different from that of the calibration film. In this case, the measured dose will be incorrect. This paper proposed a film calibration method that includes the effect of PIT.Methods:: The dose versus film optical density was fitted to a power function with three parameters. One of these parameters was PIT dependent, while the other two were found to be almost constant with a standard deviation of the mean less than 4%. The PIT‐dependent parameter was fitted to another power function of PIT. The EBT2 film model was calibrated using the PDD method with 14 different PITs ranging from 1 h to 2 months. Ten of the fourteen PITs were used for finding the fitting parameters, and the other four were used for testing the model.Results:: The verification test shows that the differences between the delivered doses and the film doses calculated with this modeling were mainly within 2% for delivered doses above 60 cGy, and the total uncertainties were generally under 5%. The errors and total uncertainties of film dose calculation were independent of the PIT using the proposed calibration procedure. However, the fitting uncertainty increased with decreasing dose or PIT, but stayed below 1.3% for this study.Conclusions:: The EBT2 film dose can be modeled as a function of PIT. For the ease of routine calibration, five PITs were suggested to be used. It is recommended that two PITs be located in the fast developing period (1∼6 h), one in 1 ∼ 2 days, one around a week, and one around a month.Keywords
Funding Information
- National Science Council (NSC 101-2221-E-214 -019)
- National Science Council (NSC 101-2221-E-214 -019)
This publication has 46 references indexed in Scilit:
- Dosimetric accuracy of Gafchromic EBT2 and EBT3 film for in vivo dosimetryAustralasian Physics & Engineering Sciences in Medicine, 2013
- Calibration of EBT2 film by the PDD method with scanner non-uniformity correctionPhysics in Medicine & Biology, 2012
- Energy dependence and dose response of Gafchromic EBT2 film over a wide range of photon, electron, and proton beam energiesMedical Physics, 2010
- Evaluation and optimization of the new EBT2 radiochromic film dosimetry system for patient dose verification in radiotherapyPhysics in Medicine & Biology, 2010
- Technical Note: Homogeneity of Gafchromic®EBT2 filmMedical Physics, 2010
- Monte Carlo calculated absorbed‐dose energy dependence of EBT and EBT2 filmMedical Physics, 2010
- Fractionated stereotactic radiotherapy: A method to evaluate geometric and dosimetric uncertainties using radiochromic filmsMedical Physics, 2009
- Optimization of the Gafchromic™ EBT protocol for IMRT QAZeitschrift für Medizinische Physik, 2009
- Film dosimetry for intensity modulated radiation therapy: Dosimetric evaluationMedical Physics, 2002
- A filtration method for improving film dosimetry in photon radiation therapyMedical Physics, 1997