Deep Learning Strategy for Sliding ECG Analysis during Cardiopulmonary Resuscitation: Influence of the Hands-Off Time on Accuracy
Open Access
Abstract
This study aims to present a novel deep learning algorithm for a sliding shock advisory decision during cardiopulmonary resuscitation (CPR) and its performance evaluation as a function of the cumulative hands-off time. We retrospectively used 13,570 CPR episodes from out-of-hospital cardiac arrest (OHCA) interventions reviewed in a period of interest from 30 s before to 10 s after regular analysis of automated external defibrillators (AEDs). Three convolutional neural networks (CNNs) with raw ECG input (duration of 5, 10, and 15 s) were applied for the shock advisory decision during CPR in 26 sequential analyses shifted by 1 s. The start and stop of chest compressions (CC) can occur at arbitrary times in sequential slides; therefore, the sliding hands-off time (sHOT) quantifies the cumulative CC-free portion of the analyzed ECG. An independent test with CPR episodes in 393 ventricular fibrillations (VF), 177 normal sinus rhythms (NSR), 1848 other non-shockable rhythms (ONR), and 3979 asystoles (ASYS) showed a substantial improvement of VF sensitivity when increasing the analysis duration from 5 s to 10 s. Specificity was not dependent on the ECG analysis duration. The 10 s CNN model presented the best performance: 92–94.4% (VF), 92.2–94% (ASYS), 96–97% (ONR), and 98.2–99.5% (NSR) for sliding decision times during CPR; 98–99% (VF), 98.2–99.8% (ASYS), 98.8–99.1 (ONR), and 100% (NSR) for sliding decision times after end of CPR. We identified the importance of sHOT as a reliable predictor of performance, accounting for the minimal sHOT interval of 2–3 s that provides a reliable rhythm detection satisfying the American Heart Association (AHA) standards for AED rhythm analysis. The presented technology for sliding shock advisory decision during CPR achieved substantial performance improvement in short hands-off periods (>2 s), such as insufflations or pre-shock pauses. The performance was competitive despite 1–2.8% point lower ASYS detection during CPR than the standard requirement (95%) for non-noisy ECG signals. The presented deep learning strategy is a basis for improved CPR practices involving both continuous CC and CC with insufflations, associated with minimal CC interruptions for reconfirmation of non-shockable rhythms (minimum hands-off time) and early treatment of VF (minimal pre-shock pauses).Keywords
Funding Information
- Bulgarian National Science Fund (KП-06-H42/3)
This publication has 53 references indexed in Scilit:
- Feasibility of automated rhythm assessment in chest compression pauses during cardiopulmonary resuscitationResuscitation, 2013
- European Resuscitation Council Guidelines for Resuscitation 2010 Section 2. Adult basic life support and use of automated external defibrillatorsResuscitation, 2010
- Minimizing pre- and post-defibrillation pauses increases the likelihood of return of spontaneous circulation (ROSC)Resuscitation, 2010
- Cardiopulmonary resuscitation artefact suppression using a Kalman filter and the frequency of chest compressions as the reference signalResuscitation, 2010
- Shock Advisory System for Heart Rhythm Analysis During Cardiopulmonary Resuscitation Using a Single ECG Input of Automated External DefibrillatorsAnnals of Biomedical Engineering, 2010
- Reduction of CPR artifacts in the ventricular fibrillation ECG by coherent line removalBioMedical Engineering OnLine, 2010
- Techniques for artefact filtering from chest compression corrupted ECG signals: Good, but not enoughResuscitation, 2009
- A method to remove CPR artefacts from human ECG using only the recorded ECGResuscitation, 2008
- Identifying potentially shockable rhythms without interrupting cardiopulmonary resuscitation*Critical Care Medicine, 2008
- Feasibility of shock advice analysis during CPR through removal of CPR artefacts from the human ECGResuscitation, 2004