Comparison of the Complior Analyse device with Sphygmocor and Complior SP for pulse wave velocity and central pressure assessment
Authors:
- Erwan BozecEBErwan Bozec
- Sandrine MillasseauSMSandrine Millasseau
- Hakim KhettabHKHakim Khettab
- Pierre BoutouyriePBPierre Boutouyrie
- Stéphane LaurentSLStéphane Laurent
- 1 April 2014
- journal article
- research article
- Published by Wolters Kluwer Health in Journal Of Hypertension
- Vol. 32 (4) , 873-880
Abstract
The Complior device (Alam Medical, France) was used in epidemiological studies which established pulse wave velocity (PWV) as a cardiovascular risk marker. Central pressure is related, but complementary to PWV and also associated to cardiovascular outcomes. The new Complior Analyse measures both PWV and central blood pressure during the same acquisition. The aim of this study was to compare PWV values from Complior Analyse with the previous Complior SP (PWVcs) and with Sphygmocor (PWVscr; AtCor, Australia), and to compare central systolic pressure from Complior Analyse and Sphygmocor. Peripheral and central pressures and PWV were measured with the three devices in 112 patients. PWV measurements from Complior Analyse were analysed using two foot-detection algorithms (PWVca_it and PWVca_cs). Both radial (ao-SBPscr) and carotid (car-SBPscr) approaches from Sphygmocor were compared to carotid Complior Analyse measurements (car-SBPca). The same distance and same calibrating pressures were used for all devices. PWVca_it was strongly correlated to PWVscr (R(2) = 0.93, P < 0.001) with a difference of 0.0 ± 0.7 m/s. PWVca_cs was also correlated to PWVcs (R(2) = 0.90, P < 0.001) with a difference of 0.1 ± 0.7 m/s. Central systolic pressures were strongly correlated. The difference between car-SBPca and ao-SBPscr was 3.1 ± 4.2 mmHg (P < 0.001), statistically equivalent to the difference between car-SBPscr and ao-SBPscr (3.9 ± 5.8 mmHg, P < 0.001), whilst the difference between car-SBPca and car-SBPscr was negligible (-0.7 ± 5.6 mmHg, P = NS). The new Complior Analyse device provides equivalent results for PWV and central pressure values to the Sphygmocor and Complior SP. It reaches Association for the Advancement of Medical Instrumentation standard for central blood pressure and grades as excellent for PWV on the Artery Society criteria. It can be interchanged with existing devices.Keywords
This publication has 44 references indexed in Scilit:
- Measurement accuracy of non-invasively obtained central blood pressure by applanation tonometry: A systematic review and meta-analysisInternational Journal of Cardiology, 2013
- Large Artery Stiffening and Remodeling Are Independently Associated With All-Cause Mortality and Cardiovascular Events in Chronic Kidney DiseaseHypertension, 2012
- Expert consensus document on the measurement of aortic stiffness in daily practice using carotid-femoral pulse wave velocityJournal Of Hypertension, 2012
- Prognostic Value of Aortic Stiffness and Calcification for Cardiovascular Events and Mortality in Dialysis PatientsClinical Journal of the American Society of Nephrology, 2011
- Determinants of pulse wave velocity in healthy people and in the presence of cardiovascular risk factors: ‘establishing normal and reference values’European Heart Journal, 2010
- Wave Reflection and Arterial Stiffness in the Prediction of 15-Year All-Cause and Cardiovascular MortalitiesHypertension, 2010
- Plasma Fatty Acid Composition as a Predictor of Arterial Stiffness and MortalityHypertension, 2009
- Stiffness of Capacitive and Conduit ArteriesHypertension, 2005
- Aortic Stiffness Is an Independent Predictor of Fatal Stroke in Essential HypertensionStroke, 2003
- Determination of pulse wave velocities with computerized algorithmsAmerican Heart Journal, 1991