Comparison of five algorithms for the detection of ventricular fibrillation from the surface ECG
Author:
- 1 November 2000
- journal article
- research article
- Published by IOP Publishing in Physiological Measurement
- Vol. 21 (4) , 429-439
Abstract
The introduction and widening application of automatic external defibrillators (AEDs) present very strong requirements for external ECG signal analysis. Highly accurate discrimination between shockable and non-shockable rhythms is required, with sensitivity and specificity aimed to approach the maximum values of 100%. We undertook an assessment of the performance of five detection algorithms, selected from among several others for their good published results. Test signals were 71 8 s ECG episodes on sinus rhythm and 90 8 s episodes on ventricular fibrillation, which were taken from the well known ECG-signal databases of the American Heart Association (AHA) and the Massachusetts Institute of Technology-Beth Israel Hospital (MIT-BIH - `cudb' and `vfdb' files). The purpose of this study is to assess the accuracy of the algorithms with signals other than those used for their development. An expected reduction of the sensitivity and specificity was found. The results could be used for further assessment, e.g. of noise and artefact sensitivity, for comparison with newly developed algorithms, etc.Keywords
This publication has 12 references indexed in Scilit:
- Linear and non-linear analysis of the surface electrocardiogram during human ventricular fibrillation shows evidence of order in the underlying mechanism.Medical & Biological Engineering & Computing, 1999
- Automatic External Defibrillators for Public Access Defibrillation: Recommendations for Specifying and Reporting Arrhythmia Analysis Algorithm Performance, Incorporating New Waveforms, and Enhancing SafetyCirculation, 1997
- ASSESSMENT OF METROLOGICAL CHARACTERISTICS OF DIGITAL ELECTROCARDIOGRAPHSJournal of Clinical Engineering, 1996
- Multiway sequential hypothesis testing for tachyarrhythmia discriminationIEEE Transactions on Biomedical Engineering, 1994
- Comparison of four techniques for recognition of ventricular fibrillation from the surface ECGMedical & Biological Engineering & Computing, 1993
- Can the techniques of nonlinear dynamics detect chaotic behavior in electrocardiographic signals?Journal of Electrocardiology, 1991
- Is fibrillation chaos?Circulation Research, 1990
- Ventricular fibrillation detection by autocorrelation function peak analysisJournal of Electrocardiology, 1990
- Algorithmic sequential decision-making in the frequency domain for life threatening ventricular arrhythmias and imitative artefacts: a diagnostic systemJournal of Biomedical Engineering, 1989