Anterior Chamber Angle Measurement with Anterior Segment Optical Coherence Tomography: A Comparison between Slit Lamp OCT and Visante OCT
Open Access
- 1 August 2008
- journal article
- research article
- Published by Association for Research in Vision and Ophthalmology (ARVO) in Investigative Ophthalmology & Visual Science
- Vol. 49 (8) , 3469-3474
Abstract
Purpose. To compare anterior chamber angle measurements obtained from two anterior segment optical coherence tomography (OCT) instruments and to evaluate their agreements and interobserver reproducibility. methods. Forty-nine eyes from 49 healthy normal subjects were studied. The anterior chamber angle was imaged with the Visante anterior segment OCT (Carl Zeiss Meditec, Dublin, CA) and the slit lamp OCT (SLOCT, Heidelberg Engineering, GmbH, Dossenheim, Germany) on one randomly selected eye in each subject and measured by two independent observers. The angle-opening distance (AOD 500), the trabecular–iris angle (TIA 500), and the trabecular–iris space area (TISA 500) at the nasal and temporal angles were measured. The agreements between SLOCT and Visante OCT measurements and the interobserver reproducibility were evaluated. results. The mean nasal/temporal anterior chamber angles measured by Visante OCT and SLOCT were 527 ± 249/572 ± 275 μm (AOD), 0.180 ± 0.091/0.193 ± 0.102 mm2 (TISA), and 38.1 ± 12.3/39.6 ± 13.2° (TIA); and 534 ± 234/628 ± 254 μm (AOD), 0.191 ± 0.089/0.217 ± 0.093 mm2(TISA), and 37.8 ± 10.1/40.6 ± 10.7° (TIA), respectively. No significant difference was found between Visante OCT and SLOCT measurements except the temporal TISA (P = 0.034). The interobserver coefficient of variation ranged between 4.4% and 7.8% for Visante OCT and 4.9% and 7.0% for SLOCT. The spans of 95% limits of agreement of the nasal/temporal angle measurements between Visante OCT and SLOCT were 437/531 mm2, 0.174/0.186 mm2, and 25.3/28.0° for AOD, TISA, and TIA, respectively. conclusions. Although Visante OCT and SLOCT demonstrate high interobserver reproducibility for anterior chamber angle measurements, their agreement was poor.Keywords
This publication has 9 references indexed in Scilit:
- Repeatability and reproducibility of anterior chamber angle measurement with anterior segment optical coherence tomographyBritish Journal of Ophthalmology, 2007
- Anterior chamber angle measurement with optical coherence tomography: Intraobserver and interobserver variabilityJournal of Cataract & Refractive Surgery, 2006
- Comparison of Optical Coherence Tomography and Ultrasound Biomicroscopy for Detection of Narrow Anterior Chamber AnglesArchives of Ophthalmology (1950), 2005
- Noncontact Goniometry With Optical Coherence TomographyArchives of Ophthalmology (1950), 2005
- Static and dynamic analysis of the anterior segment with optical coherence tomographyJournal of Cataract & Refractive Surgery, 2004
- Group index of the human cornea at 13-µm wavelength obtained in vitro by optical coherence domain reflectometryOptics Letters, 2004
- Ultrasound biomicroscopic quantitative analysis of light-dark changes associated with pupillary blockAmerican Journal of Ophthalmology, 1999
- Ultrasound biomicroscopy. II. Intraobserver and interobserver reproducibility of measurementsActa Ophthalmologica Scandinavica, 1998
- Ultrasound Biomicroscopy of Anterior Segment Structures in Normal and Glaucomatous EyesAmerican Journal of Ophthalmology, 1992