Closed-Loop Control of a Multifunctional Myoelectric Prosthesis With Full-State Anatomically Congruent Electrotactile Feedback
Open Access
- 14 April 2023
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Neural Systems and Rehabilitation Engineering
- Vol. 31 (15344320) , 2090-2100
Abstract
State-of-the-art myoelectric hand prostheses provide multi-functional control but lack somatosensory feedback. To accommodate the full functionality of a dexterous prosthesis, the artificial sensory feedback needs to convey several degrees of freedom (DoF) simultaneously. However, this is a challenge with current methods as they are characterized by a low information bandwidth. In this study, we leverage the flexibility of a recently developed system for simultaneous electrotactile stimulation and electromyography (EMG) recording to present the first solution for closed-loop myoelectric control of a multifunctional prosthesis with full-state anatomically congruent electrotactile feedback. The novel feedback scheme (coupled encoding) conveyed proprioceptive (hand aperture, wrist rotation) and exteroceptive information (grasping force). The coupled encoding was compared to the conventional approach (sectorized encoding) and incidental feedback in 10 able-bodied and one amputee participant who used the system to perform a functional task. The results showed that both feedback approaches increased the accuracy of position control compared to incidental feedback. However, the feedback increased completion time, and it did not significantly improve grasping force control. Importantly, the performance of the coupled feedback was not significantly different compared to the conventional scheme, despite the latter being easier to learn during training. Overall, the results indicate that the developed feedback can improve prosthesis control across multiple DoFs but they also highlight the subjects’ ability to exploit minimal incidental information. Importantly, the current setup is the first to convey three feedback variables simultaneously using electrotactile stimulation while providing multi-DoF myoelectric control with all hardware components mounted on the same forearm.Keywords
Funding Information
- Independent Research Fund Denmark (8022-00243A (ROBIN), 8022-00226B)
- European Union H2020 Program through the Project TACTILITY (H2020-ICT-25-856718)
This publication has 56 references indexed in Scilit:
- An exploration of grip force regulation with a low-impedance myoelectric prosthesis featuring referred haptic feedbackJournal of NeuroEngineering and Rehabilitation, 2015
- Surveying the interest of individuals with upper limb loss in novel prosthetic control techniquesJournal of NeuroEngineering and Rehabilitation, 2015
- Vibrotactile grasping force and hand aperture feedback for myoelectric forearm prosthesis usersProsthetics and Orthotics International, 2015
- Closed-Loop Control of Grasping With a Myoelectric Hand Prosthesis: Which Are the Relevant Feedback Variables for Force Control?IEEE Transactions on Neural Systems and Rehabilitation Engineering, 2014
- Linear and Nonlinear Regression Techniques for Simultaneous and Proportional Myoelectric ControlIEEE Transactions on Neural Systems and Rehabilitation Engineering, 2014
- A comparison of the real-time controllability of pattern recognition to conventional myoelectric control for discrete and simultaneous movementsJournal of NeuroEngineering and Rehabilitation, 2014
- Visuomotor behaviours when using a myoelectric prosthesisJournal of NeuroEngineering and Rehabilitation, 2014
- Stiffness Feedback for Myoelectric Forearm Prostheses Using Vibrotactile StimulationIEEE Transactions on Neural Systems and Rehabilitation Engineering, 2013
- The role of feed-forward and feedback processes for closed-loop prosthesis controlJournal of NeuroEngineering and Rehabilitation, 2011
- A robust, real-time control scheme for multifunction myoelectric controlIEEE Transactions on Biomedical Engineering, 2003