Intraindividual Variability in Serial Measurements of D l CO and Alveolar Volume over One Year in Eight Healthy Subjects Using Three Independent Measuring Systems
Authors:
- Donald P. TashkinDTDonald P. Tashkin
- Michael S. SimmonsMSMichael S. Simmons
- 1 December 1989
- journal article
- research article
- Published by Oxford University Press (OUP) in American Review of Respiratory Disease
- Vol. 140 (6) , 1818-1822
Abstract
To evaluate intraindividual and between-machine variability in Dl CO, one technician performed triplicate measurement of Dl CO in eight healthy subjects on three consecutive days using three different systems (Collins [C], Gould [G], SRL [S]) on five separate occasions over the course of 1 yr. For each subject and machine, we calculated the coefficient of variation (CV) of the Dl CO and its associated alveolar volume (Va) and inspiratory vital capacity (IVC) over the five different measurement periods using the mean value for each period. CV (in %) for Dl CO in each of the eight subjects for each machine varied from 2.4 to 7.1% for machine C, 4.1 to 10.5% for machine G, and 2.5 to 8.7% for machine S. Two-way ANOVA showed significant differences in CV for Dl CO and IVC but not for Va across subjects (p < 0.01) and for Dl CO, Va, and IVC across machines (p < 0.03): CV for Dl CO was higher for G than for C or S; CV for Va and IVC was lower for C than for G or S. Multiple regression did not show a significant relationship between within-individual CV for Dl CO and that for Va or IVC. We conclude that variability in Dl CO measurements is affected by both subject and machine, but, at least in healthy cooperative subjects, it does not appear to be affected by variability in Va or IVC. For the average subject and the two less variable machines we tested, Dl CO measurements could vary within the same subject as much as ± 9.6% (two-tailed 95% confidence interval) with repeated measurements over 1 yr.Keywords
This publication has 8 references indexed in Scilit:
- Diurnal Variation of the Diffusing Capacity of the Lung: Is it Real?American Review of Respiratory Disease, 1987
- Effect of ethanol on transfer factor: the importance of posture.Thorax, 1987
- Effect of Low Hemoglobin Levels on the Diffusing Capacity of the Lungs for CORespiration, 1985
- The effect of abnormal levels of hematocrit on the single breath diffusing capacityLung, 1982
- Attachment for Automated Single Breath Diffusing Capacity MeasurementChest, 1973
- Iron-Deficiency Anaemia: Its Effect on Transfer Factor for the Lung (Diffusing Capacity) and Ventilation and Cardiac Frequency during Sub-Maximal ExerciseClinical Science, 1972
- A THEORETICAL AND EXPERIMENTAL ANALYSIS OF ANOMALIES IN THE ESTIMATION OF PULMONARY DIFFUSING CAPACITY BY THE SINGLE BREATH METHODQuarterly Journal of Experimental Physiology and Cognate Medical Sciences, 1961
- THE ABSORPTION OF CARBON MONOXIDE BY THE LUNGS DURING BREATHHOLDING 1Journal of Clinical Investigation, 1954