Urinary Tract Infection Following Instrumentation for Urodynamic Testing
Authors:
- Charles E. WrightCWCharles E. Wright
- Nita AndresNANita Andres
- Maureen A. KozaMKMaureen A. Koza
- 1 January 1989
- journal article
- Published by Cambridge University Press (CUP) in Infection Control
- Vol. 10 (01) , 26-32
Abstract
After identifying a temporal cluster of urinary tract infections in patients who had undergone urodynamic procedures, we examined the techniques within the urodynamic laboratory and retrospectively reviewed charts of all 155 patients tested in the previous six months. The rate of acquired urinary tract infections was 18.7%. Risk factors for infection included undergoing cystometrograms and being subject to the first procedure performed in a day. Technical errors in the performance of the urodynamic studies included failure to completely disassemble the apparatus upon completion of a procedure, failure to use sterile components, and lapses in aseptic technique. Bacteria implicated in the outbreak were isolated from tubing, transducers, and flush solutions. After the institution of appropriate technique, all patients tested in the subsequent six months were followed. The rate of acquired urinary tract infection dropped to 5%. Urodynamic apparatus should be completely disassembled following the completion of a procedure; reassembly using sterile components should occur immediately prior to the next procedure; aseptic technique should be maintained; and patients should undergo routine urine screening before a procedure. Surveillance of urodynamic procedures may reveal correctable flaws in technique.Keywords
This publication has 7 references indexed in Scilit:
- Urinary tract pressure recording equipment as a source of infectionJournal of Hospital Infection, 1985
- Posterior Urethral Valve Masquerading as the Prune Belly SyndromeBritish Journal of Urology, 1983
- Bacterial Contamination of Arterial LinesJAMA, 1983
- Endemic rate of fluid contamination and related septicemia in arterial pressure monitoringThe American Journal of Medicine, 1981
- Diagnostic Catheterisation and Bacteriuria in Women with Urinary IncontinenceBritish Journal of Urology, 1978
- The Inhibitory Effect of Coli-Endotoxin on Alpha-Adrenergic Receptor Functions in the Lower Urinary TractScandinavian Journal of Urology and Nephrology, 1977
- SPASTIC STRIATED EXTERNAL SPHINCTER AND URINARY TRACT INFECTION IN GIRLS1British Journal of Urology, 1971