Double lock-in detection for recovering weak coherent radio frequency signals
Author:
- 1 August 1985
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 56 (8) , 1662-1664
Abstract
A single radio frequency lock-in amplifier reduces broadband noise, but not rf pickup of the same frequency as the signal. If this pickup noise is at least 14 dB stronger than broadband noise, after both have passed through the lock-in, then the signal-to-noise ratio can be improved by applying the lock-in output to a second, low-frequency lock-in which is synchronized to an independent modulation of the signal. Weak coherent radio frequency signals buried in both rf pickup and broadband noise can be recovered by using this double lock-in method, as demonstrated in a plasma diagnostics experiment.Keywords
This publication has 1 reference indexed in Scilit:
- Far-infrared laser scattering in the ACT-I toroidal deviceJournal of Vacuum Science & Technology A, 1985