Long trace profiler downward- and sideways-facing measurement simulations
Open Access
- 1 January 2026
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 97 (1)
Abstract
Maintaining optimal x-ray beam quality in accelerator-based light-sources requires mirrors with exceptional shape and surface precision. In practice, mirrors are often mounted in diverse orientations—such as in Kirkpatrick–Baez systems—where gravitational and mechanical effects may differ significantly from upward-facing metrology configurations. In this study, we use optical ray tracing to investigate alternative metrology geometries suitable for measuring side-facing and downward-facing mirrors. We propose a modified profilometry approach for use with the European Synchrotron Radiation Facility long trace profiler, in which beam folding optics are integrated beneath the scanning head to redirect the measurement beam. The approach combines a detailed optical model of the instrument implemented using ray-tracing, enabling accurate evaluation of the impact of motion-induced errors under realistic acquisition conditions. For downward-facing configurations, a promising solution was found experimentally, then characterized, and investigated with an analysis of the impact of alignment tolerance. While preliminary insights are provided for the side-facing setup, all proposed configurations exhibited sensitivity to motion errors of the scanning head, highlighting challenges for robust implementation.Keywords
Funding Information
- Agence Nationale de la Recherche (21-LCV1-0013-01)
This publication has 10 references indexed in Scilit:
- Towards new generation long trace profiler LTP-2020: system design with different sensors in different operation modes2023
- Towards a new generation long trace profiler LTP-2020: optical design of pencil beam interferometry sensor2023
- Raytracing the long trace profiler2020
- The Optical Metrology Laboratory at Diamond: pushing the limits of nano-metrology2019
- Stitching methods at the European Synchrotron Radiation Facility (ESRF)Review of Scientific Instruments, 2016
- Optimal alignment of mirror-based pentaprisms for scanning deflectometric devicesOptical Engineering, 2011
- ESRF metrology laboratory: overview of instrumentation, measurement techniques, and data analysis2010
- Modern Developments in X-Ray and Neutron Optics2008
- Optimal use of pentaprisms in highly accurate deflectometric scanningMeasurement Science and Technology, 2006
- Mirror metrology and bender characterization at ESRF2005