A phantom study comparing radial trajectories for accelerated cardiac 4D flow MRI against a particle imaging velocimetry reference
- 5 February 2021
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 86 (1) , 363-371
Abstract
Purpose: Radial sampling is one method to accelerate 4D flow MRI acquisition, making feasible dual‐velocity encoding (Venc) assessment of slow flow in the left ventricle (LV). Here, two radial trajectories are compared in vitro for this application: 3D radial (phase‐contrast vastly undersampled isotropic projection, PC‐VIPR) versus stack of stars (phase‐contrast stack of stars, PC‐SOS), with benchtop particle imaging velocimetry (PIV) serving as a reference standard.Methods: The study contained three steps: (1) Construction of an MRI‐ and PIV‐compatible LV model from a healthy adult’s CT images. (2) In vitro PIV using a pulsatile flow pump. (3) In vitro dual‐Venc 4D flow MRI using PC‐VIPR and PC‐SOS (two repeat experiments). Each MR image set was retrospectively undersampled to five effective scan durations and compared with the PIV reference. The root‐mean‐square velocity vector difference (RMSE) between MRI and PIV images was compared, along with kinetic energy (KE) and wall shear stress (WSS).Results: RMSE increased as scan time decreased for both MR acquisitions. RMSE was 3% lower in PC‐SOS images than PC‐VIPR images in 30‐min scans (3.8 vs. 3.9 cm/s) but 98% higher in 2.5‐min scans (9.5 vs. 4.8 cm/s). PIV intrasession repeatability showed a RMSE of 4.4 cm/s, reflecting beat‐to‐beat flow variation, while MRI had intersession RMSEs of 3.8/3.5 cm/s for VIPR/SOS, respectively. Speed, KE, and WSS were overestimated voxel‐wise in 30‐min MRI scans relative to PIV by 0.4/0.3 cm/s, 0.2/0.1 μJ/mL, and 36/43 mPa, respectively, for VIPR/SOS.Conclusions: PIV is feasible for application‐specific 4D flow MRI protocol optimization. PC‐VIPR is better‐suited to dual‐Venc LV imaging with short scan times.Keywords
Funding Information
- GE Healthcare
This publication has 37 references indexed in Scilit:
- Left ventricular thrombus formation after acute myocardial infarctionHeart, 2012
- Quantification of left and right ventricular kinetic energy using four-dimensional intracardiac magnetic resonance imaging flow measurementsAmerican Journal of Physiology-Heart and Circulatory Physiology, 2012
- Four‐dimensional phase contrast MRI with accelerated dual velocity encodingJournal of Magnetic Resonance Imaging, 2012
- Quantification and visualization of cardiovascular 4D velocity mapping accelerated with parallel imaging or k-t BLAST: head to head comparison and validation at 1.5 T and 3 TJournal of Cardiovascular Magnetic Resonance, 2011
- Hepatic vascular flow measurements by phase contrast MRI and doppler echography: A comparative and reproducibility studyJournal of Magnetic Resonance Imaging, 2010
- Improved 3D phase contrast MRI with off‐resonance corrected dual echo VIPRMagnetic Resonance in Medicine, 2008
- Quantitative 2D and 3D phase contrast MRI: Optimized analysis of blood flow and vessel wall parametersMagnetic Resonance in Medicine, 2008
- Resampling of data between arbitrary grids using convolution interpolationIEEE Transactions on Medical Imaging, 1999
- Effects of intravoxel velocity distributions on the complex difference method of phase‐contrast MR angiographyJournal of Magnetic Resonance Imaging, 1996
- Three‐Point Phase‐Contrast Velocity Measurements with Increased Velocity‐to‐Noise RatioMagnetic Resonance in Medicine, 1995