Sustained release vancomycin-coated titanium alloy using a novel electrostatic dry powder coating technique may be a potential strategy to reduce implant-related infection
Authors:
- 1 January 2017
- journal article
- Published by International Research and Cooperation Association for Bio & Socio-Sciences Advancement (IRCA-BSSA) in BioScience Trends
- Vol. 11 (3) , 346-354
Abstract
In order to tackle the implant-related infection, a novel way was developed in this study to coat vancomycin particles mixed with controlled release coating materials onto the surface of titanium alloy by using an electrostatic dry powder coating technique. To characterize this sustained release antibacterial coating, surface morphology, in vitro and in vivo drug release were sequentially evaluated. In vitro cytotoxicity was tested by Cell Counting Kit-8 (CCK-8) assay and cytological changes were observed by inverted microscope. The antibacterial properties against MRSA, including a bacterial growth inhibition assay and a colony-counting test by spread plate method were performed. Results indicated that the vancomycin-coated sample was biocompatible for Human osteoblast cell line MG-63 and displayed effective antibacterial ability against MRSA. The coating film was revealed uniform by scanning electron microscopy. Both the in vitro and in vivo drug release kinetics showed an initially high release rate, followed by an extended period of sustained drug release over 7 days. These results suggest that with good biocompatibility and antibacterial ability, the sustained release antibacterial coating of titanium alloy using our novel electrostatic dry powder coating process may provide a promising candidate for the treatment of orthopedic implant-related infection.Keywords
This publication has 20 references indexed in Scilit:
- Management of Periprosthetic Joint Infection: The Current KnowledgeJournal of Bone and Joint Surgery, 2012
- Sustained release coating of tablets with Eudragit® RS/RL using a novel electrostatic dry powder coating processInternational Journal of Pharmaceutics, 2010
- The Socioeconomic Impact of Musculoskeletal InfectionsJournal of Bone and Joint Surgery, 2010
- Incidence of Prosthetic Joint Infections After Primary Knee ArthroplastyThe Journal of Arthroplasty, 2010
- Periprosthetic Joint Infection: The Incidence, Timing, and Predisposing FactorsClinical Orthopaedics and Related Research, 2008
- Efficacy of antibiotic‐impregnated cement in total hip replacementActa Orthopaedica, 2008
- The incidence of deep prosthetic infections in a specialist orthopaedic hospitalThe Journal of Bone and Joint Surgery. British volume, 2006
- Surface modification of titanium, titanium alloys, and related materials for biomedical applicationsMaterials Science and Engineering: R: Reports, 2004
- Histomorphometric, ultrastructural and microhardness evaluation of the osseointegration of a nanostructured titanium oxide coating by metal-organic chemical vapour deposition: an in vivo studyBiomaterials, 2004
- Surface studies of in vitro biocompatibility of titanium oxide coatingsApplied Surface Science, 2001