Improved atomic force microscopy resolution using an electric double layer
Authors:
- 17 February 1997
- journal article
- research article
- Published by AIP Publishing in Applied Physics Letters
- Vol. 70 (7) , 844-846
Abstract
High resolution (“atomic”) images of clinochlore and muscovite have been obtained in aqueous solution by inducing an electric double layer between the atomic force microscope tip and the sample surface. The electric double layer is created by the addition of a surfactant to water and greatly improves image resolution. A theoretical model is proposed to explain the improved resolution.Keywords
This publication has 9 references indexed in Scilit:
- Molecular Organization of Surfactants at Solid-Liquid InterfacesScience, 1995
- Direct Visualization of Surfactant Hemimicelles by Force Microscopy of the Electrical Double LayerLangmuir, 1994
- On the limits of the spectroscopic ability of AFM and the interaction between an AFM tip and a sampleSurface Science, 1994
- Atomic Force Microscopy: Imaging with Electrical Double Layer InteractionsLangmuir, 1994
- Manipulation of van der Waals forces to improve image resolution in atomic-force microscopyJournal of Applied Physics, 1993
- A new force sensor incorporating force-feedback control for interfacial force microscopyReview of Scientific Instruments, 1991
- Forces in atomic force microscopy in air and waterApplied Physics Letters, 1989
- Force dependences for the definition of the atomic force microscopy spatial resolutionPhysics Letters A, 1988
- The Lifshitz-Van der Waals constantJournal of Colloid and Interface Science, 1972