Non-invasive measurement of internal diameter of peripheral arteries during the cardiac cycle

Authors:
  • JEJean-Daniel Etienne
    Jean-Daniel Etienne
  • PFPierre-André Farine
    Pierre-André Farine
  • PMPatrick Monney
    Patrick Monney
  • FPFrançois Perret
    François Perret
  • MCMarco Cecchini
    Marco Cecchini
  • EGEric Gagnebin
    Eric Gagnebin
  • CBClaude Bornoz
    Claude Bornoz
  • YTYannick Tardy
    Yannick Tardy
  • MAMarcel Arditi
    Marcel Arditi
  • JMJean-Jacques Meister
    Jean-Jacques Meister
  • CLClaude-Eric Leuenberger
    Claude-Eric Leuenberger

Abstract
An ultrasonic device was developed for continuous monitoring of the internal diameter of human peripheral arteries. The diastolic internal diameter of the radial artery measured in 10 supine normotensive subjects averaged 2.36 ± 0.12 mm (mean ± s.e.) and the change in diameter during the pulse wave was 0.037 ± 0.003 mm. It was surprising to find that when the transducer was used to apply pressure on the radial artery, the difference between the systolic and the diastolic diameter was significantly increased while the diastolic diameter was reduced. This method therefore allows continuous, accurate and non-invasive measurement of the internal diameter of the human radial artery.