KLT-based quality controlled compression of single-lead ECG
Authors:
- G.C. KemberGKG.C. Kember
- G.A. FentonGFG.A. Fenton
- 1 July 1998
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 45 (7) , 942-945
Abstract
An electrocardiogram (ECG) compression algorithm based on a combination of the Karhunen-Loeve transform (KLT) and multirate sampling is introduced. The use of multirate sampling reduces KLT computational times to those reported for wavelet-packet-based compression techniques. A beat-by-beat quality controlled compression criterion is shown to be necessary to ensure clinically adequate reconstruction of each beat. The resulting quality controlled algorithm efficiently achieves compression rates of approximately 30-40:1 for the MIT-BIH database.Keywords
This publication has 4 references indexed in Scilit:
- Wavelet packet-based compression of single lead ECGIEEE Transactions on Biomedical Engineering, 1996
- Multichannel ECG data compression by multirate signal processing and transform domain coding techniquesIEEE Transactions on Biomedical Engineering, 1993
- Multirate digital filters, filter banks, polyphase networks, and applications: a tutorialProceedings of the IEEE, 1990
- Orthogonal Transforms for Digital Signal Processing1975