The performance of liquid chromatographic columns has been significantly improved as a result of the development of packings made with a controlled surface porosity (CSP) support. The regular nature of the packing minimizes trans- column variations in flow, and the thin film of sorbent used on the support provides for minimal stationary phase mass transfer limitations. The H.E.T.P. of liquid-liquid chromatographic columns made from this support show only moderate increase with increasing carrier velocity; therefore, separations can be carried out at high carrier velocities with only a slight decrease in column efficiency. Columns prepared with this support have demonstrated more than 2,600 theoretical plates per foot for unretained solutes. For rapidly eluting compounds, more than 40 theoretical plates per second can be obtained with 2–3 mm., i.d., columns operated at carrier velocities of about 2 cm/sec. Excellent separation of isomers and homologs may be effected in a few minutes using these conditions. Critical parameters in the use of the CSP support have been studied. The factors investigated include: level of stationary phase loading (liquid film thickness); support types; support surface porosity; column internal diameter; and reproducibility of column preparation. One of the practical advantages of the CSP support is the ease and convenience in preparing reproducible, high-efficiency liquid-liquid chromatographic columns. Packings made with support particles as small as 20 microns are free flowing and readily dry packed into columns by conventional techniques. Since the CSP support is mechanically stable, columns made with this material can be conveniently operated at high input pressures.