Characterisation of an electron collecting CdTe strip sensor using the MYTHEN readout chip
- 1 January 2015
- journal article
- Published by IOP Publishing in Journal of Instrumentation
- Vol. 10 (1) , C01024
Abstract
MYTHEN is a single photon counting hybrid strip X-ray detector that has found application in x-ray powder diffraction (XRPD) experiments at synchrotrons worldwide. Originally designed to operate with hole collecting silicon sensors, MYTHEN is suited for detecting X-rays above 5 keV, however many PD beamlines have been designed for energies above 50 keV where silicon sensors have an efficiency of only few percent. In order to adapt MYTHEN to meet these energies the absorption efficiency of the sensor must be substantially increased. Cadmium-Telluride (CdTe) has an absorption efficiency approximately 30 times that of silicon at 50 keV, and is therefore a very promising replacement candidate for silicon. Furthermore, the large dynamic range of the pre-amplifier of MYTHEN and its double polarity capability has enabled the characterisation of an electron collecting Schottky type CdTe sensor. A CdTe MYTHEN system has undergone a series of characterisation experiments including stress test of bias and radiation induced polarizations. The performance of this system will be presented and discussed.Keywords
This publication has 10 references indexed in Scilit:
- The Materials Science beamline upgrade at the Swiss Light SourceJournal of Synchrotron Radiation, 2013
- Characterization of polarization phenomenon in Al-Schottky CdTe detectors using a spectroscopic analysis methodNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2011
- The MYTHEN detector for X-ray powder diffraction experiments at the Swiss Light SourceJournal of Synchrotron Radiation, 2010
- MythenII: A 128 channel single photon counting readout chipNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2009
- The polarization mechanism in CdTe Schottky detectorsApplied Physics Letters, 2009
- A mechanism for dynamic lateral polarization in CdZnTe under high flux x-ray irradiationApplied Physics Letters, 2008
- Characteristics of both carriers with polarization in diode-type CdTe x-ray detectorsApplied Physics Letters, 2007
- Performance of a new Schottky CdTe detector for hard X-ray spectroscopy2002
- Temperature dependence of the band gap energy of crystalline CdTeJournal of Physics and Chemistry of Solids, 2000
- Radiation induced polarization in CdTe detectorsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1988