In situ generated PANI promoted flexible photoelectrochemical biosensor for ochratoxin A based on GOx-stuffed DNA hydrogel as enhancer
- 28 February 2023
- journal article
- research article
- Published by Springer Nature in Microchimica Acta
- Vol. 190 (3) , 1-10
Abstract
A flexible photoelectrochemical (PEC) biosensor is proposed for the sensitive detection of ochratoxin A (OTA) based on glucose oxidase (GOx)-encapsulated target-responsive hydrogel, using Fenton reaction-mediated in situ formation of polyaniline (PANI) as signal amplified strategy. The target-responsive DNA hydrogels with high loading capacity can carry a large amount of GOx, which not only avoids laborious labeling process but also enhances the analytical performance. Upon introduction of target molecules, the hydrogel can be opened, and multiple GOx was released, thus producing lots of H2O2 via catalytic reduction of glucose. As a component of the Fenton reagent, H2O2 can react with the Fe2+ on the graphene oxidase-PAMAM-Fe2+ (GO-PAMAM-Fe2+) to generate Fe3+ and ·OH. This in turn can oxidize aniline and generate polyaniline (PANI), resulting in the enhancement of the photocurrent signal of GO-MoS2-CdS photoelectrode. The GO-PAMAM-Fe2+ as the neighborhood component of GO-MoS2-CdS-based photoactive material not only can increase the loading amount of Fe2+, but also can inhibit the decrease of photocurrent of GO-MoS2-CdS by direct modification of Fe2+ on the photoactive material. Moreover, the high loading capacity of DNA hydrogel can efficiently promote the performance of the PEC biosensor. The PEC biosensor exhibited satisfactory analytical performance for OTA with a linear range of 0.0001–0.1 ng/mL and a low detection limit of 0.05 pg/mL. It presents recommendable specificity, stability, and practical applications. Importantly, the PEC biosensor provides a new concept for construction of PEC biosensing platform. Graphical abstractKeywords
Funding Information
- National Natural Science Foundation of China (21864013, 21904056)
- Natural Science Foundation of Jiangxi Province (20202ZDB01003, 20212BCJL23057, 20202BAB203008)
This publication has 41 references indexed in Scilit:
- Amplified impedimetric immunosensor based on instant catalyst for sensitive determination of ochratoxin ABiosensors and Bioelectronics, 2016
- A fluorescent aptasensor based on DNA-scaffolded silver nanoclusters coupling with Zn(II)-ion signal-enhancement for simultaneous detection of OTA and AFB1Sensors and Actuators B: Chemical, 2016
- Silver Nanolabels-Assisted Ion-Exchange Reaction with CdTe Quantum Dots Mediated Exciton Trapping for Signal-On Photoelectrochemical Immunoassay of MycotoxinsAnalytical Chemistry, 2016
- Gold Nanoparticles Deposited Polyaniline–TiO2 Nanotube for Surface Plasmon Resonance Enhanced Photoelectrochemical BiosensingACS Applied Materials & Interfaces, 2015
- Insight into the Effect of Highly Dispersed MoS2 versus Layer-Structured MoS2 on the Photocorrosion and Photoactivity of CdS in Graphene–CdS–MoS2 CompositesThe Journal of Physical Chemistry C, 2015
- Dual-functional MoS2 sheet-modified CdS branch-like heterostructures with enhanced photostability and photocatalytic activityJournal of Materials Chemistry A, 2013
- Target-Responsive “Sweet” Hydrogel with Glucometer Readout for Portable and Quantitative Detection of Non-Glucose TargetsJournal of the American Chemical Society, 2013
- Multiplexed Fluorescence Resonance Energy Transfer Aptasensor between Upconversion Nanoparticles and Graphene Oxide for the Simultaneous Determination of MycotoxinsAnalytical Chemistry, 2012
- A novel hydrogen peroxide biosensor based on enzymatically induced deposition of polyaniline on the functionalized graphene–carbon nanotube hybrid materialsSensors and Actuators B: Chemical, 2011
- Novel highly-performing immunosensor-based strategy for ochratoxin A detection in wine samplesBiosensors and Bioelectronics, 2008