Abstract
Rationale: The World Anti‐Doping Agency (WADA) encourages drug‐testing laboratories to develop screening methods that can detect as many doping substances as possible in urine. The use of full‐scan high‐resolution acquisition (FS/HR) with gas chromatography/mass spectrometry (GC/MS) for the detection of known and unknown trimethylsilyl (TMS) derivatives of anabolic‐androgenic steroids (AAS) provides anti‐doping testing bodies with a new analytical tool. Methods: The AAS were extracted from urine samples by generic liquid–liquid extraction, after enzymatic hydrolysis, and TMS derivatization. The extracted urine was analyzed by GC/Q‐TOF and GC/Q‐Orbitrap to compare the performance of the two instrument types for the detection of 46 AAS in human urine. The quantitation of endogenous anabolic steroids and the ability of the two analytical platforms to comply with the requirements for testing as part of the WADA Athlete Biological Passport (ABP) were also assessed. Results: The data presented show that the analytical performance for both instruments complies with the WADA specifications. The limits of detection (LODs) for both instruments are well below the WADA 50% Minimum Required Performance Levels. The mass errors in the current study for the GC/Q‐Orbitrap platform are lower than those obtained for the GC/Q‐TOF instrument. Conclusions: The data presented herein proved that both molecular profiling platforms can be used for antidoping screening. The mass accuracies are excellent in both instruments; however, the GC/Q‐Orbitrap performs better as it provides higher resolution than the GC/Q‐TOF platform.Keywords
Funding Information
- Qatar National Research Fund (NPRP 6 - 334 - 3 - 087, NPRP 6-334-3-087)
This publication has 22 references indexed in Scilit:
- Gas chromatographic quadrupole time-of-flight full scan high resolution mass spectrometric screening of human urine in antidoping analysisJournal of Chromatography B, 2017
- Non‐targeted acquisition strategy for screening doping compounds based on GC‐EI‐hybrid quadrupole‐Orbitrap mass spectrometry: A focus on exogenous anabolic steroidsDrug Testing and Analysis, 2017
- Quality classification of Spanish olive oils by untargeted gas chromatography coupled to hybrid quadrupole-time of flight mass spectrometry with atmospheric pressure chemical ionization and metabolomics-based statistical approachFood Chemistry, 2017
- Analysis of volatile organic compounds released from the decay of surrogate human models simulating victims of collapsed buildings by thermal desorption–comprehensive two-dimensional gas chromatography–time of flight mass spectrometryAnalytica Chimica Acta, 2015
- Comprehensive two‐dimensional gas chromatography‐mass spectrometry: Recent evolution and current trendsMass Spectrometry Reviews, 2014
- An improved gas chromatography screening method for doping substances using triple quadrupole mass spectrometry, with an emphasis on quality assuranceBiomedical Chromatography, 2012
- Gas chromatography coupled to high-resolution time-of-flight mass spectrometry to analyze trace-level organic compounds in the environment, food safety and toxicologyTrAC Trends in Analytical Chemistry, 2011
- Analytical challenges in doping control: Comprehensive two-dimensional gas chromatography with time of flight mass spectrometry, a promising optionJournal of Chromatography A, 2009
- Preventive doping control analysis: liquid and gas chromatography time‐of‐flight mass spectrometry for detection of designer steroidsRapid Communications in Mass Spectrometry, 2007
- Mass spectrometry in sports drug testing: Structure characterization and analytical assaysMass Spectrometry Reviews, 2006