Interference between PET and MRI sub-systems in a silicon-photomultiplier-based PET/MRI system
Authors:
- Seiichi YamamotoSYSeiichi Yamamoto
- Hiroshi WatabeHWHiroshi Watabe
- Yasukazu KanaiYKYasukazu Kanai
- Masaaki AokiMAMasaaki Aoki
- Eiji SugiyamaESEiji Sugiyama
- Tadashi WatabeTWTadashi Watabe
- Masao ImaizumiMIMasao Imaizumi
- Eku ShimosegawaESEku Shimosegawa
- Jun HatazawaJHJun Hatazawa
- 21 June 2011
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 56 (13) , 4147-4159
Abstract
The silicon-photomultiplier (Si-PM) is a promising photodetector, especially for integrated PET/MRI systems, due to its small size, high gain, and low sensitivity to static magnetic fields. The major problem using a Si-PM-based PET system within the MRI system is the interference between the PET and MRI units. We measured the interference by combining a Si-PM-based PET system with a permanent-magnet MRI system. When the RF signal-induced pulse height exceeded the lower energy threshold level of the PET system, interference between the Si-PM-based PET system and MRI system was detected. The prompt as well as the delayed coincidence count rates of the Si-PM-based PET system increased significantly. These noise counts produced severe artifacts on the reconstructed images of the Si-PM-based PET system. In terms of the effect of the Si-PM-based PET system on the MRI system, although no susceptibility artifact was observed on the MR images, electronic noise from the PET detector ring was detected by the RF coil and reduced the signal-to-noise ratio (S/N) of the MR images. The S/N degradation of the MR images was reduced when the distance between the RF coil and the Si-PM-based PET system was increased. We conclude that reducing the interference between the PET and MRI systems is essential for achieving the optimum performance of integrated Si-PM PET/MRI systems.Keywords
This publication has 24 references indexed in Scilit:
- Evaluation of Geiger-mode APDs for PET block detector designsPhysics in Medicine & Biology, 2010
- Avoiding the ballistocardiogram (BCG) artifact of EEG data acquired simultaneously with fMRI by pulse-triggered presentation of stimuliJournal of Neuroscience Methods, 2010
- Features of Silicon Photo Multipliers: Precision Measurements of Noise, Cross-Talk, Afterpulsing, Detection EfficiencyIEEE Transactions on Nuclear Science, 2009
- An Investigation Into the Use of Geiger-Mode Solid-State Photomultipliers for Simultaneous PET and MRI AcquisitionIEEE Transactions on Nuclear Science, 2008
- Simultaneous PET-MRI: a new approach for functional and morphological imagingNature Medicine, 2008
- Simultaneous in vivo positron emission tomography and magnetic resonance imagingProceedings of the National Academy of Sciences of the United States of America, 2008
- Complete artifact removal for EEG recorded during continuous fMRI using independent component analysisNeuroImage, 2007
- Ultimate intrinsic signal‐to‐noise ratio in MRIMagnetic Resonance in Medicine, 1998
- Minimizing interference from magnetic resonance imagers during electrocardiographyIEEE Transactions on Biomedical Engineering, 1998
- PET and NMR dual acquisition (PANDA): applications to isolated, perfused rat heartsNMR in Biomedicine, 1997