QA of intensity-modulated beams using dynamic MLC log files
Authors:
Open Access
- 1 January 2006
- journal article
- research article
- Published by Medknow in Journal of Medical Physics
- Vol. 31 (1) , 36-41
Abstract
To evaluate the utility of Dynalog file information for planar dose verification in IMRT QA, a program is developed to convert Dynalog file data to DMLC field files. For this study, five predefined fluencies are planned and delivered using Varian, Eclipse 3D planning system and 6MV photon beam of Varian, Clinac DMX linear accelerator. To measure planar dose distribution, Kodak, EDR2 films are exposed in similar setup as planning setup. Dynalog files are recorded for each delivery and converted into DMLC field files using in-house program. Delivered dose distributions are calculated using DMLC field files from Dynalog files. Planned, Measured and Delivered dose distributions are compared using gamma evaluation in Scanditronix, Omni Pro IMRT software. The Planned and Delivered planar dose distributions agree within 2% dose difference and 2 mm DTA. Measured dose distributions agree within 4% dose difference and 4 mm DTA with Planned dose distribution. Our results show Dynalog file as a promising tool for dynamic IMRT QA.Keywords
This publication has 5 references indexed in Scilit:
- Evaluation of the gamma dose distribution comparison methodMedical Physics, 2003
- Validation of dynamic MLC-controller log files using a two-dimensional diode arrayMedical Physics, 2003
- Verification of dynamic and segmental IMRT delivery by dynamic log file analysisJournal of Applied Clinical Medical Physics, 2002
- Comprehensive quality assurance for the delivery of intensity modulated radiotherapy with a multileaf collimator used in the dynamic modeMedical Physics, 2001
- A technique for the quantitative evaluation of dose distributionsMedical Physics, 1998