Thermal effects generated by high-intensity focused ultrasound beams at normal incidence to a bone surface
Authors:
- 1 January 2010
- journal article
- Published by Acoustical Society of America (ASA) in The Journal of the Acoustical Society of America
- Vol. 127 (1) , 549-559
Abstract
Experiments and computations were performed to study factors affecting thermal safety when high-intensity focused ultrasound (HIFU) beams are normally incident (i.e., beam axis normal to the interface) upon a bone/soft-tissue interface. In particular, the temperature rise and thermal dose were determined as a function of separation between the beam focus and the interface. Under conditions representative of clinical HIFU procedures, it was found that the thermal dose at the bone surface can exceed the threshold for necrosis even when the beam focus is more than 4 cm from the bone. Experiments showed that reflection of the HIFU beam from the bone back into the transducer introduced temperature fluctuations of as much as ±15% and may be an important consideration for safety analyses at sufficiently high acoustic power. The applicability of linear propagation models in predicting thermal dose near the interface was also addressed. Linear models, while underpredicting thermal dose at the focus, provided a conservative (slight overprediction) estimate of thermal dose at the bone surface. Finally, temperature rise due to absorption of shear waves generated by the HIFU beam in the bone was computed. Modeling shear-wave propagation in the thermal analysis showed that the predicted temperature rise off axis was as much as 30% higher when absorption of shear waves is included, indicating that enhanced heating due to shear-wave absorption is potentially important, even for normally incident HIFU beams.Keywords
This publication has 32 references indexed in Scilit:
- Palliative Treatment of Painful Bone Metastases with MR Imaging–guided Focused UltrasoundRadiology, 2008
- Investigation of the viscous heating artefact arising from the use of thermocouples in a focused ultrasound fieldPhysics in Medicine & Biology, 2008
- High-Intensity Focused Ultrasound Therapy: an Overview for RadiologistsKorean Journal of Radiology, 2008
- Ultrasound-Induced Thermal Elevation in Clotted Blood and Cranial BoneUltrasound in Medicine & Biology, 2007
- HIFU procedures at moderate intensities—effect of large blood vesselsPhysics in Medicine & Biology, 2007
- Prediction of temperature rise in layered media from measured ultrasonic intensity dataPhysics in Medicine & Biology, 1994
- Ultrasonic heating of the skullThe Journal of the Acoustical Society of America, 1990
- Solutions of the bio-heat transfer equationPhysics in Medicine & Biology, 1988
- Thermal conduction errors of manganin-constantan thermocouple arraysPhysics in Medicine & Biology, 1985
- Compilation of empirical ultrasonic properties of mammalian tissues. IIThe Journal of the Acoustical Society of America, 1980