Effects of some technological aspects on the fatigue strength of a cementless hip stem
Authors:
- 1 July 1995
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research
- Vol. 29 (7) , 875-881
Abstract
Four prototype cementless hip stems were tested following the ISO 7206 protocol for the assessment of the endurance properties and compared with a cast CrCoMo (ASTM F75) commercially available stem which was used as reference design. All the tested stems were similar in shape and size, but with some substantial differences. The first was made of forged Ti6Al4V alloy (ASTM F136). The second, made of the same material, featured a central hole intended to reduce the bending stiffness of the stem itself. The third was identical to the second but for a small tooling notch in one of the fillets of the hole. The fourth was similar to the first but had a coating of sintered titanium beads in the proximal part. All of these modifications were made to evaluate the effect of fatigue strength of intentional or unintentional features commonly found in commercial stems. The forged Ti6Al4V alloy was found to be substantially stronger than the cast ASTM F75 CrCoMo alloy. However, tooling notches or sintered coatings were found to dramatically reduce this strength. Thus, the Ti6Al4V alloy calls for an accurate design process, especially when complex shapes or sintered structures are required. © 1995 John Wiley & Sons, Inc.Keywords
This publication has 12 references indexed in Scilit:
- Fracture of the femoral stem of the Ring TCH hip prosthesisThe Journal of Bone and Joint Surgery. British volume, 1992
- Testing of Total Hip Replacements: Endurance Tests and Stress Measurements: Part 1: Endurance TestsProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1990
- Long-term results of bilateral Charnley total hip arthroplastyThe Journal of Arthroplasty, 1988
- Failure of the stem in total hip replacementArchives of orthopaedic and trauma surgery, 1988
- Hüftendoprothesenschaftfrakturen - eine klinische und werkstofftechnische UntersuchungZeitschrift für Orthopädie und ihre Grenzgebiete, 1983
- Fatigue properties of implant materials in hip prosthesis form: A standardized testJournal of Biomedical Materials Research, 1983
- Fracture failure of a forged vitallium prosthesis. A case report.Journal of Bone and Joint Surgery, 1982
- Fracture of hip prostheses due to welding defectsArchives of orthopaedic and trauma surgery, 1981
- Five years experience with ceramic-metal-composite hip endoprosthesesArchives of orthopaedic and trauma surgery, 1981
- Fracture of Femoral Prostheses in Total Hip ReplacementClinical Orthopaedics and Related Research, 1975