Abstract
In this article we present a method that aims towards more ideal number concentration response of a diffusion charger and diffusion collector-based particle sensor (Dekati ePNC) by utilizing reduced pressure conditions in the sensor. A model is derived for the charger and diffusion collector stage of the sensor as well as for the total particle number concentration response. The derived model is validated with experimental characterization measurements in different operation pressures. The experimental results show that reduced operation pressure makes the sensor response less dependent on particle size, approaching pure particle number response. The obtained results improve the feasibility of the diffusion-charging based method in the measurement of particle number concentration.
Funding Information
  • Dekati Ltd., Kangasala, Finland
  • Tampere University Aerosol Measurement Infrastructure, through the Academy of Finland Infrastructure Funding (337225)

This publication has 21 references indexed in Scilit: