A Tandem Mass Spectrometer for Improved Transmission and Analysis of Large Macromolecular Assemblies
- 16 February 2002
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 74 (6) , 1402-1407
Abstract
We report the design and first applications of a tandem mass spectrometer (a quadrupole time-of-flight mass spectrometer) optimized for the transmission and analysis of large macromolecular assemblies. Careful control of the pressure gradient in the different pumping stages of the instrument has been found to be essential for the detection of macromolecular particles. Such assemblies are, however, difficult to analyze by tandem-MS approaches, because they give rise to signals above m/z 3000−4000, the limit for commercial quadrupoles. By reducing the frequency of the quadrupole to 300 kHz and using it as a narrow-band mass filter, we show that it is possible to isolate ions from a single peak at m/z 22 000 in a window as narrow as 22 m/z units. Using cesium iodide cluster signals, we show that the mass range in the time-of-flight (TOF) analyzer extends beyond m/z 90 000, in theory to more than m/z 150 000. We also demonstrate that the resolution of the instrument is greater than 3000 at m/z 44 500. Tandem-MS capabilities are illustrated by separating components from heterooligomeric assemblies formed between tetrameric transthyretin, thyroxine, retinol-binding protein, and retinol. Isolation of a single charge state at m/z 5340 in the quadrupole and subsequent collision-induced dissociation (CID) in the gas-filled collision cell leads to the formation of ions from individual subunits and subcomplexes, identified by their mass and charge in the TOF analyzer.Keywords
This publication has 17 references indexed in Scilit:
- Role of conserved residues in structure and stability: Tryptophans of human serum retinol‐binding protein, a model for the lipocalin superfamilyProtein Science, 2001
- The effect of the source pressure on the abundance of ions of noncovalent protein assemblies in an electrospray ionization orthogonal time‐of‐flight instrumentRapid Communications in Mass Spectrometry, 2001
- MALDI Quadrupole Time-of-Flight Mass Spectrometry: A Powerful Tool for Proteomic ResearchAnalytical Chemistry, 2000
- ELECTROSPRAY AS AN IONISATION METHOD FOR MASS SPECTROMETRYJournal of Aerosol Science, 1999
- Detection of the Intact GroEL Chaperonin Assembly by Mass SpectrometryJournal of the American Chemical Society, 1999
- Probing the nature of interactions in SH2 binding interfaces–evidence from electrospray ionization mass spectrometryProtein Science, 1999
- High Sensitivity Collisionally-activated Decomposition Tandem Mass Spectrometry on a Novel Quadrupole/Orthogonal-acceleration Time-of-flight Mass SpectrometerRapid Communications in Mass Spectrometry, 1996
- A high mass range quadrupole spectrometer for cluster studiesInternational Journal of Mass Spectrometry and Ion Processes, 1989
- Generation and detection of cluster ions [(CsI)nCs]+ ranging up to m/z = 90 000International Journal of Mass Spectrometry and Ion Processes, 1984
- Interference Effects in the QuasimolecularX-Ray Production Probability for 10-MeV-Ar CollisionsPhysical Review Letters, 1983