Theoretical analysis of non-invasive oscillometric maximum amplitude algorithm for estimating mean blood pressure
Authors:
- 1 May 1997
- journal article
- research article
- Published by Springer Nature in Medical & Biological Engineering & Computing
- Vol. 35 (3) , 271-278
Abstract
A theoretical analysis is performed to evaluate the effect of arterial mechanical and blood pressure pulse properties on the accuracy of non-invasive oscillometric maximum amplitude algorithm (MAA) estimates of the mean blood pressure obtained using air-filled occlusive cuffs. Invasively recorded blood pressure pulses, selected for their varied shapes, are scaled to simulate a wide range of blood pulse pressures (diastolic blood pressure minus systolic blood pressure). Each scaled blood pressure pulse is transformed through an exponential model of an artery to create a series of blood volume pulses from which a simulated oscillometric waveform is created and the corresponding MAA estimate of the mean blood pressure and error (mean blood pressure minus MAA estimate) are determined. The MAA estimates are found to depend on the arterial blood pressure. The errors are found to depend on the arterial mechanical properties, blood pressure pulse shape and blood pulse pressure. These results suggest that there is no direct relationship between the mean blood presure and MAA estimate, and that multiple variables may affect the accuracy of MAA estimates of the mean blood pressure obtained using air-filled occlusive cuffs.Keywords
This publication has 24 references indexed in Scilit:
- A mathematical study of some biomechanical factors affecting the oscillometric blood pressure measurementIEEE Transactions on Biomedical Engineering, 1996
- Changes in oscillometric pulse amplitude envelope with cuff size: implications for blood pressure measurement criteria and cuff size selectionJournal of Biomedical Engineering, 1993
- Comparative accuracies of a finger blood pressure monitor and an oscillometric blood pressure monitorJournal of Clinical Monitoring and Computing, 1991
- Accuracy of four indirect methods of blood pressure measurement, with hemodynamic correlationsJournal of Clinical Monitoring and Computing, 1990
- Forearm arterial pressure-volume relationships in manClinical Physics and Physiological Measurement, 1988
- Oscillometric Determination of Diastolic, Mean and Systolic Blood Pressure—A Numerical ModelJournal of Biomechanical Engineering, 1986
- The Measurement of Continuous Finger Arterial Pressure Noninvasively in Stationary Subjects1986
- Correlation study of arterial blood pressure level to the amplitude of the pressure pulse waveformJournal of Biomedical Engineering, 1984
- The Meaning of the Point of Maximum Oscillations in Cuff Pressure in the Indirect Measurement of Blood Pressure—Part IIJournal of Biomechanical Engineering, 1980
- Indirect Measurement of Mean Blood Pressure in the Anesthetized PatientAnesthesiology, 1979