Dose Reduction With Dedicated CT Metal Artifact Reduction Algorithm: CT Phantom Study

Authors:

Abstract
OBJECTIVE. The objective of this study was to compare reader accuracy detecting lesions near hardware in a CT phantom model at different radiation exposures using an advanced metal artifact reduction (MAR) algorithm and standard filtered back projection (FBP) techniques and to determine if radiation exposure could be decreased using MAR without compromising lesion detectability. MATERIALS AND METHODS. A CT phantom manufactured with spherical lesions of various sizes (10–20 mm) and attenuations (20–50 HU) embedded around cobalt-chromium spheres attached to titanium rods, simulating an arthroplasty, was scanned on a single CT scanner (FLASH, Siemens Healthcare) at 140 kVp and 0.6-mm collimation using clinical-dose (300 Quality Reference mAs [Siemens Healthcare]), low-dose (150 Quality Reference mAs), and high-dose (600 Quality Reference mAs) protocols. Images reconstructed with iterative MAR, advanced modeled iterative reconstruction (ADMIRE), and FBP with identical parameters were anonymized and independen...