Effects of Atmospheric Ozone on Microarray Data Quality
Authors:
- 15 July 2003
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 75 (17) , 4672-4675
Abstract
A data anomaly was observed that affected the uniformity and reproducibility of fluorescent signal across DNA microarrays. Results from experimental sets designed to identify potential causes (from microarray production to array scanning) indicated that the anomaly was linked to a batch process; further work allowed us to localize the effect to the posthybridization array stringency washes. Ozone levels were monitored and highly correlated with the batch effect. Controlled exposures of microarrays to ozone confirmed this factor as the root cause, and we present data that show susceptibility of a class of cyanine dyes (e.g., Cy5, Alexa 647) to ozone levels as low as 5−10 ppb for periods as short as 10−30 s. Other cyanine dyes (e.g., Cy3, Alexa 555) were not significantly affected until higher ozone levels (>100 ppb). To address this environmental effect, laboratory ozone levels should be kept below 2 ppb (e.g., with filters in HVAC) to achieve high quality microarray data.Keywords
This publication has 4 references indexed in Scilit:
- Expression profiling using microarrays fabricated by an ink-jet oligonucleotide synthesizerNature Biotechnology, 2001
- Drug target validation and identification of secondary drug target effects using DNA microarraysNature Medicine, 1998
- Biologically Active Cyanine Dyes as Probes for the Identification of Active Oxygen Species1992
- PHOTOBLEACHING OF A CYANINE DYE IN SOLUTION AND IN MEMBRANESPhotochemistry and Photobiology, 1987