Course of Metal Ions after a Revision of Malfunctioning Metal-on-Metal Total Hip Prostheses
Open Access
Abstract
The present research evaluated the course of cobalt and chromium in the blood and urine after the revision of metal-on-metal with a ceramic-on-polyethylene total hip arthroplasty. Seven patients were enrolled for hip prosthesis revision owing to ascertained damage of the implant. Metals in the blood and urine were evaluated before and after the hip revision. The double measurement before the total hip revision revealed high levels of metal ions (on average, 88.1 µg/L of cobalt in the blood, 399.0 µg/g of creatinine cobalt in the urine, 46.8 µg/L of chromium in the blood, and 129.6 µg/g of creatinine chromium in the urine at the first measurements), with an increasing trend between the first and second dosage. Within a week after the hip revision, the levels of metal ions significantly decreased by approximately half. Four to six months after the operation, the cobalt levels were found near to the reference values, whereas the chromium levels reached 25% of the values measured before the revision. The revision of malfunctioning metal-on-metal implants produced a dramatic decrease of metal ions in biological fluids, although it did not completely rescue the chromium level.Keywords
This publication has 33 references indexed in Scilit:
- Metal Ion Concentrations in Body Fluids after Implantation of Hip Replacements with Metal-on-Metal Bearing – Systematic Review of Clinical and Epidemiological StudiesPLOS ONE, 2013
- Cobalt metabolism and toxicology—A brief updateScience of The Total Environment, 2012
- Metal ion levels in large-diameter total hip and resurfacing hip arthroplasty-Preliminary results of a prospective five year study after two years of follow-upBMC Musculoskeletal Disorders, 2012
- The Role of Albumin in Human Toxicology of Cobalt: Contribution from a Clinical CaseISRN Hematology, 2011
- Metal Ion Release With Large-Diameter Metal-on-Metal Hip ArthroplastyThe Journal of Arthroplasty, 2011
- Polyneuropathy caused by cobalt–chromium metallosis after total hip replacementMuscle & Nerve, 2010
- Exposure to chromium, cobalt and molybdenum from metal-on-metal total hip replacement and hip resurfacing arthroplastyActa Orthopaedica, 2006
- CobaltJournal of Toxicology: Clinical Toxicology, 1999
- Distribution of Cobalt Chromium Wear and Corrosion Products and Biologic ReactionsClinical Orthopaedics and Related Research, 1996
- Health risks associated with cobalt exposure — an overviewScience of The Total Environment, 1994