An experimental comparison of different devices for pulsatile high-pressure lavage and their relevance to cement intrusion into cancellous bone
- 1 September 2007
- journal article
- research article
- Published by Springer Nature in Archives of orthopaedic and trauma surgery
- Vol. 127 (10) , 873-877
Abstract
Introduction Some of the systems marketed for pulsatile high-pressure lavage are clearly different in their mechanical pulse characteristics. To take the matter further we set up an experimental in vitro model to determine whether these different lavage systems might produce different cementation results because of their differing pulse characteristics, and whether the systems differed significantly in terms of cement depth penetration. Materials and methods A total of 48 femoral heads were obtained at operation from patients who had undergone endoprosthetic hip replacement. The specimens were subjected to manual rinsing or semiautomatic cleansing procedures with five different pulsatile lavage devices. After the cleansing procedures, polymethylmetacrylate bone cement was intruded into the cancellous bone under standardised conditions. Determination of cement penetration was done by computed tomography using an image processing software. Results Irrespective of the nature of the lavage system tested, superficial cement penetration, both up to 2 mm and up to 3 mm, was significantly better after pulsatile high-pressure lavage than it was after manual rinsing of the specimens with a bladder syringe (P < 0.001 in each case). Whereas our experimental model did not show any significant differences between the lavage systems as regards the cementation results to a depth of up to 2 mm (P = 0.996), there were significant differences at a target depth of 3 mm (P < 0.05). Conclusion As compared with manual rinsing, pulsatile high-pressure lavage in vitro makes highly significant improvements in cement penetration into cancellous bone and should be regarded as an indispensable component of modern cementation techniques for endoprosthetic surgery. Up to the present we have not been able to ascertain what influence the different pulse characteristics of a pulsatile lavage system may have on the intended cementation result, whether it is impact force, pulse shape, pulse duration, flow rate or frequency. Appropriate investigations will be necessary.Keywords
This publication has 22 references indexed in Scilit:
- The measurement and comparison of jet characteristics of surgical pulse lavage devicesThe Journal of Arthroplasty, 2003
- Die Bedeutung der Jet-Lavage für das In-vitro- und In-vivo-ZementierergebnisZeitschrift für Orthopädie und ihre Grenzgebiete, 2001
- Lavage technique in total hip arthroplastyThe Journal of Arthroplasty, 2000
- Importance of a Thin Cement MantleClinical Orthopaedics and Related Research, 1998
- Prophylaxis of Fat and Bone Marrow Embolism in Cemented Total Hip ArthroplastyClinical Orthopaedics and Related Research, 1998
- In vivo effect of pressurization of polymethyl methacrylate bone-cementThe Journal of Arthroplasty, 1995
- Suboptimal (thin) distal cement mantle thickness as a contributory factor in total hip arthroplasty femoral component failureThe Journal of Arthroplasty, 1994
- The Mechanism of Loosening of Cemented Acetabular Components in Total Hip ArthroplastyClinical Orthopaedics and Related Research, 1992
- Histologic changes in the dog femur following total hip replacement with current cementing techniquesThe Journal of Arthroplasty, 1986
- The effect of thickness and pressure on the curing of PMMA bone cement for the total hip joint replacementJournal of Biomechanics, 1980