The In Vitro Effects of Instrumentation on Multilevel Cervical Strut-Graft Mechanics
Authors:
- Denis J. DiAngeloDDDenis J. DiAngelo
- Y. Raja RampersaudYRY. Raja Rampersaud
- Keith A. VosselKVKeith A. Vossel
- Thomas H. JansenTJThomas H. Jansen
- 1 November 1999
- journal article
- research article
- Published by Wolters Kluwer Health in Spine
- Vol. 24 (22) , 2366-76
Abstract
Biomechanical study using a programmable testing apparatus that replicated physiologic flexion–extension cervical spine motion, and loading mechanics. To determine the influence of anterior, posterior, or combined plating on multilevel cervical strut-graft mechanics in vitro. The addition of instrumentation does not prevent construct failure in multilevel ( more than two levels) cervical corpectomy. Six fresh human cadaveric cervical spines (C2–T1) were tested in six sequential conditions that included harvested (H), C4–6 corpectomy, strut grafted, strut grafted with an anterior cervical plate (SGAP), strut grafted with posterior plates (SGPP), and strut grafted with combined anterior and posterior plates (SGAPP). A customized force-sensing strut graft (FSSG) was used to measure axial compression–tension, flexion–extension and lateral bending moments, and axial torsion. Parameters of stiffness, segmental vertebral motion, and strut-graft loads were compared, to determine differences among the spine conditions. Flexion of the strut-grafted spine loaded the FSSG, and extension motion unloaded the FSSG. With the anterior plate, flexion of the SGAP spine significantly unloaded the FSSG; extension loaded the FSSG more than flexion of the unplated spine (P= 0.03). The opposite occurred with the posterior plates (SGPP), where flexion of the spine significantly loaded the FSSG (more than the strut grafted spine) and extension unloaded the FSSG (P< 0.03). The combined construct (SGAPP) counteracted the tension band effect of the individual plates and demonstrated significantly less overall FSSG load change than either plate alone (P= 0.03). Multilevel cervical instrumentation effectively increases stiffness after corpectomy. However, anterior or posterior plating alone excessively loads the graft with small degrees of motion, which may promote pistoning and failure of multilevel constructs.Keywords
This publication has 55 references indexed in Scilit:
- Biomechanical evaluation of Caspar and Cervical Spine Locking Plate systems in a cadaveric modelJournal of Neurosurgery, 1996
- An In-Vitro Biomechanical Comparison of the Orosco and AO Locking Plates for Anterior Cervical Spine FixationJournal of Spinal Disorders, 1995
- Biomechanics of Cervical Spine Internal FixationSpine, 1991
- Central corpectomy for cervical spondylotic myelopathy: a consecutive series with long-term follow-up evaluationJournal of Neurosurgery, 1991
- 1989 Volvo Award in Basic Science: Device-Related Osteoporosis with Spinal InstrumentationSpine, 1989
- Anterior Plate Stabilization for Bursting Teardrop Fractures of the Cervical SpineSpine, 1988
- Cervical Stabilization by Plate and Bone FusionSpine, 1988
- Multichannel strain gauge telemetry for orthopaedic implantsJournal of Biomechanics, 1988
- Anterior Plate Fixation of Traumatic Lesions of the Lower Cervical SpineSpine, 1987
- An in-vitro study of the kinematics of the normal, injured and stabilized cervical spineJournal of Biomechanics, 1984