- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 44 (4) , 1053-1065
Abstract
Equivalent fields are often used in radiation oncology for calculation of dose. This avoids the need to make a scatter integration but has limited applicability and some inaccuracy. We have evaluated the alternative of explicit integration of phantom-scatter dose using a functional representation, sigma, of the ratio between the scatter dose and the primary dose. The independent parameters are depth and the side of the square field. The function chosen has the advantage that the integral for a right triangle is available in closed form, which simplifies the determination of the dose from phantom-scattered photons for irregular fields by summation over such triangles. This approach accounts for the influence of depth and beam quality, which the commonly used equivalent-field tables and the area-over-perimeter relation ignore. The accuracy of this procedure is determined by the accuracy of the function sigma. This has about a 1% error of total dose for high-energy x-rays. We conclude that the tables and rules can be replaced by a computer-implemented integration of the phantom-scatter dose represented by this function sigma and using sectors or right triangles. Summing the closed-form contributions from component right triangles reduces the calculation time, which is particularly desirable when many fields are employed, as for intensity-modulated techniques and inverse planning. Measurements performed on irregular MLC-shaped fields compared well with the result from calculations.Keywords
This publication has 11 references indexed in Scilit:
- Is there a need for a revised table of equivalent square fields for the determination of phantom scatter correction factors?Physics in Medicine & Biology, 1997
- Quality control of measured x-ray beam dataMedical Physics, 1997
- Theoretical Analysis of the Equivalent Square Field by Means of the Pencil Beam ModelZeitschrift für Medizinische Physik, 1997
- Analysis of central‐axis doses for high‐energy x raysMedical Physics, 1995
- Equivalent squares of irregular photon fieldsMedical Physics, 1993
- An equivalent‐square formula for head‐scatter factorsMedical Physics, 1993
- Scatter integration with right triangular fieldsMedical Physics, 1985
- A note on equivalent circles, squares, and rectanglesMedical Physics, 1982
- Automation of Radiation Treatment Planning—IV. Derivation of a Mathematical Expression for the per cent Depth Dose Surface of Cobalt 60 Beams and Visualisation of Multiple Field Dose DistributionsBritish Journal of Radiology, 1964
- A Note on Depth Doses in Fields of Irregular ShapeBritish Journal of Radiology, 1941