Influence of breast composition and interseed attenuation in dose calculations for post-implant assessment of permanent breast103Pd seed implant
Authors:
- Jean-Philippe PignolJPJean-Philippe Pignol
- Brian KellerBKBrian Keller
- Jean-François CarrierJCJean-François Carrier
- Brigitte ReniersBRBrigitte Reniers
- Frank VerhaegenFVFrank Verhaegen
- Luc BeaulieuLBLuc Beaulieu
- 29 July 2010
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 55 (16) , 4547-4561
Abstract
The impact of tissue heterogeneity and interseed attenuation is studied in post-implant evaluation of five clinical permanent breast (103)Pd seed implants using the Monte Carlo (MC) dose calculation method. Dose metrics for the target (PTV) as well as an organ at risk (skin) are used to visualize the differences between a TG43-like MC method and more accurate MC methods capable of considering the breast tissue heterogeneity as well as the interseed attenuation. PTV dose is reduced when using a breast tissue model instead of water in MC calculations while the dose to the skin is increased. Furthermore, we investigate the effect of varying the glandular/adipose proportion of the breast tissue on dose distributions. The dose to the PTV (skin) decreases (increases) with the increasing adipose proportion inside the breast. In a complete geometry and compared to a TG43-like situation, the average PTV D(90) reduction varies from 3.9% in a glandular breast to 35.5% when the breast consists entirely of adipose. The skin D(10) increases by 28.2% in an entirely adipose breast. The results of this work show the importance of an accurate and patient-dependent breast tissue model to be used in the dosimetry for this kind of low energy implant.Keywords
This publication has 31 references indexed in Scilit:
- A Monte Carlo study on the effect of seed design on the interseed attenuation in permanent prostate implantsMedical Physics, 2008
- Postimplant Dosimetry Using a Monte Carlo Dose Calculation Engine: A New Clinical StandardInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2007
- Impact of interseed attenuation and tissue composition for permanent prostate implantsMedical Physics, 2006
- Dosimetric effects of seed anisotropy and interseed attenuation for and prostate implantsMedical Physics, 2005
- Validation ofGEANT4, an object-oriented Monte Carlo toolkit, for simulations in medical physicsMedical Physics, 2004
- Geant4—a simulation toolkitNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2003
- CERR: A computational environment for radiotherapy researchMedical Physics, 2003
- CT-based dosimetry calculations for 125I prostate implantsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1999
- Comparison of x‐ray cross sections for diagnostic and therapeutic medical physicsMedical Physics, 1996
- Skin telangiectasia: The influence of radiation dose delivery parameters in the conservative management of early breast cancerClinical Oncology, 1995