Effect of injection rate on contrast‐enhanced MR angiography image quality: Modulation transfer function analysis
- 1 August 2016
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 78 (1) , 357-369
Abstract
Purpose: Contrast‐enhanced (CE)‐MRA optimization involves interactions of sequence duration, bolus timing, contrast recirculation, and both R1 relaxivity and ‐related reduction of signal. Prior data suggest superior image quality with slower gadolinium injection rates than typically used.Methods: A computer‐based model of CE‐MRA was developed, with contrast injection, physiologic, and image acquisition parameters varied over a wide gamut. Gadolinium concentration was derived using Verhoeven's model with recirculation, R1 and calculated at each time point, and modulation transfer curves used to determine injection rates, resulting in optimal resolution and image contrast for renal and carotid artery CE‐MRA. Validation was via a vessel stenosis phantom and example patients who underwent carotid CE‐MRA with low effective injection rates.Results: Optimal resolution for renal and carotid CE‐MRA is achieved with injection rates between 0.5 to 0.9 mL/s and 0.2 to 0.3 mL/s, respectively, dependent on contrast volume. Optimal image contrast requires slightly faster injection rates. Expected signal‐to‐noise ratio varies with both contrast volume and cardiac output. Simulated vessel phantom and clinical carotid CE‐MRA exams at an effective contrast injection rate of 0.4 to 0.5 mL/s demonstrate increased resolution.Conclusion: Optimal image resolution is achieved at intuitively low, effective injection rates (0.2–0.9 mL/s, dependent on imaging parameters and contrast injection volume). Magn Reson Med 78:357–369, 2017. © 2016 International Society for Magnetic Resonance in MedicineKeywords
Funding Information
- Bracco Diagnostics
This publication has 39 references indexed in Scilit:
- Optimization of the dynamic, Gd-EOB-DTPA-enhanced MRI of the liver: the effect of the injection rateActa Radiologica, 2012
- CT iterative reconstruction in image space: A phantom studyPhysica Medica, 2011
- Can MR fluoroscopic triggering technique and slow rate injection provide appropriate arterial phase images with reducing artifacts on gadoxetic acid‐DTPA (Gd‐EOB‐DTPA)‐enhanced hepatic MR imaging?Journal of Magnetic Resonance Imaging, 2010
- Evaluation of a low-dose/slow-rotating SPECT-CT systemPhysics in Medicine & Biology, 2008
- Dependence of vessel area accuracy and precision as a function of MR imaging parameters and boundary detection algorithmJournal of Magnetic Resonance Imaging, 2007
- SENSE: Sensitivity encoding for fast MRIMagnetic Resonance in Medicine, 1999
- Optimizing Three-Dimensional Gadolinium-Enhanced Magnetic Resonance AngiographyInvestigative Radiology, 1998
- The effects of time varying intravascular signal intensity and k‐space acquisition order on three‐dimensional MR angiography image qualityJournal of Magnetic Resonance Imaging, 1996
- An evaluation of CT systems from ten manufacturersBritish Journal of Radiology, 1981
- The Sensitivity Performance of the Human Eye on an Absolute Scale*Journal of the Optical Society of America, 1948