Swab Sample Transfer for Point-Of-Care Diagnostics: Characterization of Swab Types and Manual Agitation Methods
Open Access
- 2 September 2014
- journal article
- research article
- Published by Public Library of Science (PLoS) in PLOS ONE
- Vol. 9 (9) , e105786
Abstract
The global need for disease detection and control has increased effort to engineer point-of-care (POC) tests that are simple, robust, affordable, and non-instrumented. In many POC tests, sample collection involves swabbing the site (e.g., nose, skin), agitating the swab in a fluid to release the sample, and transferring the fluid to a device for analysis. Poor performance in sample transfer can reduce sensitivity and reproducibility. In this study, we compared bacterial release efficiency of seven swab types using manual-agitation methods typical of POC devices. Transfer efficiency was measured using quantitative PCR (qPCR) for Staphylococcus aureus under conditions representing a range of sampling scenarios: 1) spiking low-volume samples onto the swab, 2) submerging the swab in excess-volume samples, and 3) swabbing dried sample from a surface. Excess-volume samples gave the expected recovery for most swabs (based on tip fluid capacity); a polyurethane swab showed enhanced recovery, suggesting an ability to accumulate organisms during sampling. Dry samples led to recovery of ∼20–30% for all swabs tested, suggesting that swab structure and volume is less important when organisms are applied to the outer swab surface. Low-volume samples led to the widest range of transfer efficiencies between swab types. Rayon swabs (63 µL capacity) performed well for excess-volume samples, but showed poor recovery for low-volume samples. Nylon (100 µL) and polyester swabs (27 µL) showed intermediate recovery for low-volume and excess-volume samples. Polyurethane swabs (16 µL) showed excellent recovery for all sample types. This work demonstrates that swab transfer efficiency can be affected by swab material, structure, and fluid capacity and details of the sample. Results and quantitative analysis methods from this study will assist POC assay developers in selecting appropriate swab types and transfer methods.Keywords
This publication has 25 references indexed in Scilit:
- Comparison of CDC and sequence-based molecular typing of syphilis treponemes: tpr and arp loci are variable in multiple samples from the same patientBMC Microbiology, 2013
- Self-Collection of Foam Nasal Swabs for Respiratory Virus Detection by PCR among Immunocompetent Subjects and Hematopoietic Cell Transplant RecipientsJournal of Clinical Microbiology, 2013
- Performance of the BD GeneOhm MRSA Achromopeptidase Assay for Real-Time PCR Detection of Methicillin-Resistant Staphylococcus aureus in Nasal SpecimensJournal of Clinical Microbiology, 2011
- Better Detection of Staphylococcus aureus Nasal Carriage by Use of Nylon Flocked SwabsJournal of Clinical Microbiology, 2010
- Swab Type, Moistening, and Preenrichment for Staphylococcus aureus on Environmental SurfacesJournal of Clinical Microbiology, 2010
- Comparison of four nasal sampling methods for the detection of viral pathogens by RT-PCR—A GA2LEN projectJournal of Virological Methods, 2009
- Evaluation of the Xpert Methicillin-Resistant Staphylococcus aureus (MRSA) Assay Using the GeneXpert Real-Time PCR Platform for Rapid Detection of MRSA from Screening SpecimensJournal of Clinical Microbiology, 2008
- Effect of Swab Composition and Use of Swabs versus Swab-Containing Skim Milk-Tryptone-Glucose-Glycerol (STGG) on Culture- or PCR-Based Detection of Streptococcus pneumoniae in Simulated and Clinical Respiratory Specimens in STGG Transport MediumJournal of Clinical Microbiology, 2008
- Distribution Study ofChlamydia trachomatisSerovars among High-Risk Women in China Performed Using PCR-Restriction Fragment Length Polymorphism GenotypingJournal of Clinical Microbiology, 2007
- Nasal swabs for detection of respiratory syncytial virus RNAArchives of Disease in Childhood, 2007