Validation of a two-axis accelerometer for monitoring patient activity during blood pressure or ECG holter monitoring
Authors:
- Jean René BorderiesJBJean René Borderies
- Odile BigaignonOBOdile Bigaignon
- Pascal GuilloPGPascal Guillo
- Philippe GossePGPhilippe Gosse
- 1 December 2003
- journal article
- research article
- Published by Wolters Kluwer Health in Blood Pressure Monitoring
- Vol. 8 (6) , 229-235
Abstract
The aim of the study was to evaluate the efficiency of a position/activity monitoring system based on a dual-axis accelerometer strapped to the subject's thigh and a position sensor located within a monitor placed on the subject's belt. Twenty-six subjects wearing two monitors (one accelerometer on each thigh) were submitted to various activities and positions under the control of an observer. An analysis of each tracing was performed both manually by a technician and automatically by dedicated software before being compared with the information gathered during the study. The accelerometer allowed accurate discrimination between the standing versus the sitting and lying positions. The sitting and lying positions were correctly detected by the built-in position sensor provided the unit was firmly attached. Walking was adequately detected by the accelerometer. The activity score was well correlated with treadmill speed. Changes in position and activity were detected with a mean error of less than 3 s. The combination of an accelerometer placed on the subject's thigh and a position sensor located at the subject's waist appeared to be a suitable system for position/activity monitoring during ambulatory ECG and blood pressure monitoring.Keywords
This publication has 13 references indexed in Scilit:
- Determinants of ambulatory blood pressure response to physical activityJournal Of Hypertension, 2002
- Microprocessor-based ambulatory activity monitoring in stroke patientsMedicine & Science in Sports & Exercise, 2002
- Physical activity level is an independent predictor of the diurnal variation in blood pressureJournal Of Hypertension, 2000
- Ambulatory Physical Activity as a Determinant of Diurnal Blood Pressure VariationHypertension, 1999
- Toward a portable blood pressure recorder device equipped with an accelerometerMedical Engineering & Physics, 1999
- Automated physical activity monitoring: Validation and comparison with physiological and self‐report measuresPsychophysiology, 1993
- Automatic Sleep/Wake Identification From Wrist ActivitySLEEP, 1992
- Actigraphy: A Means of Assessing Circadian Patterns in Human ActivityChronobiology International, 1990
- Validation of a simple mechanical accelerometer (pedometer) for the estimation of walking activityEuropean Journal of Applied Physiology, 1987
- An Activity-Based Sleep Monitor System for Ambulatory UseSLEEP, 1982