Respiratory‐Resolved Isotropic 3D T 1 Mapping of the Carotid Vessel Wall With B 1 + Correction ( CARISMAT1C )
Open Access
- 4 September 2026
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 97 (1) , 133-145
Abstract
Purpose: To develop, optimize, and characterize a respiratory motion‐resolved free‐running isotropic 3D carotid vessel wall T 1 mapping technique named CARISMAT1C. Methods: An inversion‐recovery gradient‐echo free‐running pulse sequence with interleaved double flip‐angle (2FA) was implemented, and an extended‐phase‐graph dictionary was used to map the T 1 relaxation time. In a phantom, the T 1 accuracy was compared to that of a single‐flip‐angle (1FA) variant, the inversion‐recovery spin‐echo reference, and clinical routine MOLLI, while the T 1 precision of several phyllotaxis‐based radial trajectories was compared. The most precise sequence was used in 11 healthy volunteers (25 ± 3Y, 3F). A synthetic 3D gray‐blood image was reconstructed from the source images and used to delineate the internal, external, and common carotid arteries. T 1 values across artery sections were compared to those of MOLLI. Results: In the phantom, the 2FA variant had higher T 1 agreement with the IR‐SE reference than 1FA or MOLLI. The top‐to‐bottom phyllotaxis trajectory with golden‐step shuffling was the most precise and was retained for in vivo scans. End‐expiratory carotid vessel wall T 1 values (1185 ± 42 ms) were similar to those obtained with MOLLI (1162 ± 84 ms, p = 0.37) and showed less intersubject variability and high interscan repeatability. No significant differences were found across arterial segments in the vessel wall T 1 values. Conclusion: CARISMAT1C had high T 1 accuracy in phantoms and high T 1 precision in vivo in different sections of the carotid artery tree.Keywords
Funding Information
- Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung (32003B_182615)
- Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung (CRSII5_202276)
- Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung (FNS_310030_215604)
- Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung (PZ00P3_202140)
- Schweizerische Herzstiftung (FF23129)
This publication has 44 references indexed in Scilit:
- Diagnostics Accuracy of Magnetic Resonance Imaging in Detection of Atherosclerotic Plaque Characteristics in Carotid Arteries Compared to Histology: A Systematic ReviewJournal of Magnetic Resonance Imaging, 2024
- Reliability and Value of 3D Sequential QUantitative T1‐T2‐T2* MAppings (SQUMA) MR Multi‐Parametric Imaging in Characterizing Carotid Artery AtherosclerosisJournal of Magnetic Resonance Imaging, 2022
- Three‐dimensional quantitative T1 and T2 mapping of the carotid artery: Sequence design and in vivo feasibilityMagnetic Resonance in Medicine, 2015
- Reducing contrast contamination in radial turbo‐spin‐echo acquisitions by combining a narrow‐band KWIC filter with parallel imagingMagnetic Resonance in Medicine, 2014
- Golden‐angle radial sparse parallel MRI: Combination of compressed sensing, parallel imaging, and golden‐angle radial sampling for fast and flexible dynamic volumetric MRIMagnetic Resonance in Medicine, 2013
- Spatially resolved extended phase graphs: Modeling and design of multipulse sequences with parallel transmissionMagnetic Resonance in Medicine, 2012
- Spiral phyllotaxis: The natural way to construct a 3D radial trajectory in MRIMagnetic Resonance in Medicine, 2011
- Sparse MRI: The application of compressed sensing for rapid MR imagingMagnetic Resonance in Medicine, 2007
- Swallowing, arterial pulsation, and breathing induce motion artifacts in carotid artery MRIJournal of Magnetic Resonance Imaging, 2006
- Modified Look‐Locker inversion recovery (MOLLI) for high‐resolution T1 mapping of the heartMagnetic Resonance in Medicine, 2004