Correcting for stage error motions in radius measurements
- 1 October 2005
- journal article
- Published by Optica Publishing Group in Applied Optics
- Vol. 44 (28) , 5884-5893
Abstract
Traceable radius of curvature measurements are critical for precision optics manufacturing. An optical bench measurement is repeatable and is the preferred method for low-uncertainty applications. With an optical bench, the displacement of the optic is measured as it is moved between the cat’s eye and the confocal positions, each identified using a figure measuring interferometer. The translated distance is nominally the radius of curvature; however, errors in the motion of the stage add a bias to the measurement, even if the error motions are zero on average. Estimating the bias and resulting measurement uncertainty is challenging. We have developed a new mathematical definition of the radius measurand that intrinsically corrects for error motion biases and also provides a means of representing other terms such as figure error correction, wave-front aberration biases, displacement gauge calibration and their uncertainties. With this formalism, it is no long necessary to design a high-quality radius bench to carry out a precision measurement; rather a lower quality is adequate, provided that error motions are repeatable and characterized and error motion measurement uncertainties are estimated.Keywords
This publication has 2 references indexed in Scilit:
- Displacement Uncertainty in Interferometric Radius MeasurementsCIRP Annals, 2002
- Radius measurement by interferometryOptical Engineering, 1992