Abstract
Objective. To evaluate the acceptability of the analytical characteristics of the Snibe system for the reference values of the Beckman Coulter system and in the low-concentration range. Materials and methods. Total prostate-specific antigen (PSA) concentration was measured using the Snibe analytical system (MAGLUMI 2000 Plus automated immunochemiluminescence analyzer; MAGLUMI Total PSA (CLIA) in-vitro reagent kit for quantitative determination of total PSA by immunochemiluminescent analysis on MAGLUMI automated analyzers; WHO 96/670 calibrator) and the Beckman Coulter analytical system (Access 2 automated immunochemiluminescence analyzer; Access Hybritech PSA reagent kit; Access Hybritech PSA calibrators). In the first phase of the study, blood serum samples, the three-level control material Lyphochek Tumor Marker Plus Control (Bio-Rad), and control samples from the Bio-Rad External Quality Assurance Services (EQAS) program were used to assess precision and relative accuracy. The standard measurement uncertainty (u) was assumed to be equal to the coefficient of variation under EQAS reproducibility conditions, consistent with the within-laboratory long-term inter-series coefficient of variation. The material of the second phase of the study was blood serum samples from 93 men. Blood was collected between 7.00 and 10.00 a.m. on an empty stomach. Serum was separated by centrifugation at 3000 rpm for 10 minutes. The concentration was measured in primary vacuum tubes on the day of blood collection. Statistical analysis was performed using "Statistica 10.0" software (license AXA009K287210FAACD-B). Statistical significance was set at p 0.05. When describing the data obtained, arithmetic mean (M), standard deviation (SD), median (Me), minimum and maximum values were indicated. To compare results obtained with the two analytical systems, the following methods were used: Pearson correlation coefficient (r) to assess the strength and direction of linear association, mean coefficient of variation as a relative measure of difference between paired measurements; Passing-Bablok regression analysis to identify constant and proportional bias, Bland-Altman consistency analysis to quantify agreement in clinically relevant concentration ranges; and a 2x2 contingency table to assess the effect of bias on clinical interpretation. Results. In the clinically relevant range of 1–10 ng/mL, presenting results as a concentration interval that incorporates the expanded measurement uncertainty of both analytical systems ensures their comparability when assessed against the population reference interval ( 4.0 ng/mL). However, for individual monitoring and decision-making based on specific thresholds, the identified bias must be taken into account. The Snibe-WHO analytical system shows low precision when measuring trace amounts of PSA (up to 141 % under repeatability conditions), which limits the possibilities of laboratory monitoring of patients after radical prostatectomy. Conclusions. The data obtained in our laboratory served as the basis for establishing the standard measurement uncertainty for BC-Hybr as 9.4 % in the concentration range 0.1 ng/mL (conventionally underestimated, according to repeatability precision data) and increasing the lower limit of quantitation for Snibe-WHO to 0.1 ng/mL.