Assessing the Validity of the Omron HEM‐907XL Oscillometric Blood Pressure Measurement Device in a National Survey Environment
Open Access
- 28 December 2009
- journal article
- research article
- Published by Wiley in The Journal of Clinical Hypertension
- Vol. 12 (1) , 22-28
Abstract
J Clin Hypertens (Greenwich). 2010;12:22–28. ©2009 Wiley Periodicals, Inc.Blood pressure (BP) readings taken by Omron HEM‐907XL were compared with the results obtained using sphygmomanometer (HgS) in 509 individuals using 2002 Association for the Advancement of Medical Instrumentation (AAMI) criteria. With the exception of diastolic BP in youth ages 13 to 19 years (mean difference, −1.77 mm Hg; standard deviation, 8.65), the Omron device met the criteria. Agreement for hypertension (BP ≥140/90 mm Hg) was above chance (κ=0.68) and, compared with HgS, Omron underestimated the prevalence of hypertension by 2.65%. The Omron and HgS measurements were highly correlated (r=0.94 for systolic BP and r=0.83 for diastolic BP). Both increased systolic and diastolic BP decreased device agreement (β‐coefficient=−0.10872, P<.0001; β‐coefficient=−0.25981, P<.0001, respectively). The Omron device meets AAMI criteria with the exception of diastolic BP in youth ages 13 to 19 years. However, Omron underestimated the prevalence of hypertension and device agreement decreases with increased systolic and diastolic BP.Keywords
This publication has 11 references indexed in Scilit:
- Comparison Between an Automated and Manual Sphygmomanometer in a Population SurveyAmerican Journal of Hypertension, 2008
- Validation of the Omron M5-I, R5-I and HEM-907 automated blood pressure monitors in elderly individuals according to the International Protocol of the European Society of HypertensionBlood Pressure Monitoring, 2007
- US demographic trends in mid-arm circumference and recommended blood pressure cuffs for children and adolescents: data from the National Health and Nutrition Examination Survey 1988–2004Blood Pressure Monitoring, 2007
- Comparison of Dinamap PRO-100 and Mercury Sphygmomanometer Blood Pressure Measurements in a Population-Based StudyAmerican Journal of Hypertension, 2006
- US demographic trends in mid-arm circumference and recommended blood pressure cuffs: 1988–2002Journal of Human Hypertension, 2005
- National health and nutrition examination survey 1999-2000: effect of observer training and protocol standardization on reducing blood pressure measurement errorJournal of Clinical Epidemiology, 2003
- The Seventh Report of the Joint National Committee on Prevention, Detection, Evaluation, and Treatment of High Blood PressureThe JNC 7 ReportJAMA, 2003
- Measuring Blood Pressure AccuratelyJAMA, 2003
- Validation of the Omron HEM-907 device for blood pressure measurementBlood Pressure Monitoring, 2002
- Evaluation of the overall efficacy of the Omron office digital blood pressure HEM-907 monitor in adultsBlood Pressure Monitoring, 2001