A supra-annular malposition of the Perceval S sutureless aortic valve: the ‘χ-movement’ removal technique and subsequent reimplantation
Open Access
- 25 April 2012
- journal article
- video audio-media
- Published by Oxford University Press (OUP) in Interactive CardioVascular and Thoracic Surgery
- Vol. 15 (2) , 280-281
Abstract
The Perceval S sutureless valve prosthesis has recently been introduced as a new biological aortic valve prosthesis, but a specific learning curve is required, as for every cardiac surgical centre dealing with a new technique. After the removal of the stenotic valve, the prosthetic valve is correctly positioned within the mildly decalcified aortic annulus. When a supra-annular malposition occurs, due to an excessively rapid release of the prosthesis in the aorta or incomplete annular visualization, the Perceval S valve can safely be removed even after balloon dilation. The procedure performed is a ‘χ-movement’ with the aid of anatomical forceps. If the prosthesis does not show any malformation after the procedure, it can be reimplanted in the correct intra-annular position.Keywords
This publication has 4 references indexed in Scilit:
- Sutureless valve implantation in a patient with bicuspid aortic valveInternational Journal of Cardiology, 2012
- Perceval sutureless approach in a patient with porcelain aorta unsuitable for transcatheter aortic valve implantationInternational Journal of Cardiology, 2012
- Effect of sutureless implantation of the Perceval S aortic valve bioprosthesis on intraoperative and early postoperative outcomesThe Journal of Thoracic and Cardiovascular Surgery, 2011
- Sutureless Perceval S aortic valve replacement: a multicenter, prospective pilot trial.2009