Performance of automated urine analyzers using flow cytometric and digital image-based technology in routine urinalysis
- 28 August 2019
- journal article
- research article
- Published by Taylor & Francis in Scandinavian Journal of Clinical and Laboratory Investigation
- Vol. 79 (7) , 468-474
Abstract
The purpose of this study was to evaluate the analytical performances of Sysmex UF-5000 and Dirui FUS-200 and to compare the results with manual microscopy and between each other. Two hundred fifty urine samples were analyzed for evaluation. Mid-stream specimens were studied sequentially using Dirui FUS-200 and Sysmex UF-5000, and also with manual microscopy within one hour. The physical and chemical components of urinalysis, and sediment results were investigated. The precision results of the FUS-200 and UF-5000 for WBCs, RBCs, and ECs were acceptable. The both analyzers demonstrated good linearity (r > 0.97), with no carry-over. The comparisons of FUS-200 and UF-5000 with manual microscopy for RBCs, WBCs, and ECs on 250 samples exhibited good agreement with little bias (R > 0.780). Only, the moderate agreements were obtained for calcium oxalate for both analyzers (R = 0.512, and 0.648, respectively). The sensitivities of the FUS-200 and UF-5000 were 75.8% and 86.8%, with specificities of 92.3% and 87.8% for WBCs, for RBCs the sensitivities were 91.1%, and 84.4% with specificities of 82.2%, and 89.6% for both analyzers. Kappa values of the UF-5000 were higher than FUS-200 for WBCs, RBCs, ECs, and calcium oxalate. The FUS-200 and UF-5000 urine analyzers, are both accurate, very precise systems and can be safely used in clinical laboratories. However, due to the technological characteristics of the UF-5000 analyzer, its positive impacts on the morphologic recognition and enumeration of RBCs and WBCs should be taken into account, particularly in university hospital laboratories with high patient volumes.Keywords
This publication has 17 references indexed in Scilit:
- The comparison of automated urine analyzers with manual microscopic examination for urinalysis automated urine analyzers and manual urinalysisPractical Laboratory Medicine, 2016
- Performance Evaluation and Comparison of the Fully Automated Urinalysis Analyzers UX-2000 and Cobas 6500Laboratory Medicine, 2016
- Screening of presumptive urinary tract infections by the automated urine sediment analyser sediMAXClinical Chemistry and Laboratory Medicine (CCLM), 2015
- Urine microscopy – an important diagnostic toolTidsskrift for Den norske legeforening, 2014
- Cutoff values for bacteria and leukocytes for urine sediment analyzer FUS200 in culture-positive urinary-tract infectionsScandinavian Journal of Clinical and Laboratory Investigation, 2014
- Comparison of Fully Automated Urine Sediment Analyzers H800‐FUS100 and Labumat‐Urised with Manual MicroscopyJournal of Clinical Laboratory Analysis, 2013
- Evaluation of the Sysmex UF1000i flow cytometer for ruling out bacterial urinary tract infectionClinica Chimica Acta, 2010
- Urine sediment analysis: Analytical and diagnostic performance of sediMAX® — A new automated microscopy image-based urine sediment analyserClinica Chimica Acta, 2010
- Urine sediment examination: A comparison between the manual method and the iQ200 automated urine microscopy analyzerClinica Chimica Acta, 2005
- A Coefficient of Agreement for Nominal ScalesEducational and Psychological Measurement, 1960